I/O RANGE MODIFICATION

Configurator Software is used to change input types and 
ranges.
For changing output ranges, set the Output Range Selector 
and then select by Configurator Software.

 

SIDE VIEW

Output Range Selector

V

I

I

V

I

Output Range Selector

OUTPUT RANGE

SW POSITION

V2Z1: -10 – +10 V / 0 – 20 mA DC

I

V2V2: -10 – +10 V / -10 – +10 V DC

V

STATUS INDICATOR LED

The unit is provided with a status indicator LED which 
blinks in different patterns indicating various status of its 
CPU.  When it is functioning within normal parameters, the  
LED blinks in a regular pattern of ON and OFF.
The following figure indicates typical patterns.

LED PATTERNS

STATUS

Normal

Programming Mode
Out. 1 Calibration Mode
Out. 2 Calibration Mode

Burnout

100 ms

ON

OFF

MAINTENANCE

Regular calibration procedure is explained below:

 

CALIBRATION

Warm up the unit for at least 10 minutes.  Apply 0%, 25%, 
50%, 75% and 100% input signal.  Check that the output 
signal for the respective input signal remains within accu-
racy described in the data sheet.  When the output is out of 
tolerance, recalibrate the unit according to the “CALIBRA-
TION” explained earlier.

LIGHTNING SURGE PROTECTION

M-System offers a series of lightning surge protector for 
protection against induced lightning surges.  Please contact 
M-System to choose appropriate models.