LEADWIRE RESISTANCE COMPENSATION

Unbalances in resistance between RTD wires can be com-
pensated as explained below:

 

HOW TO COMPENSATE LEADWIRE RESISTANCE

Leadwire resistance compensation can be done with the 
ZERO/SPAN Selector and UP/DOWN switches.
Turn the power supply on and warm up for more than 10 
minutes.

1) Short across the tips of RTD sensor.
2) Slide the Zero/Span Selector to the position Z while press-

ing DOWN switch.  Confirm the LED pattern showing 

“Leadwire Resistance Compensation Mode” and release 

the DOWN switch.

3) Press DOWN switch to execute the compensation.  Con-

firm that the LED turns off when you press DOWN 

switch.

4) Put the Zero/Span Selector back to the position OFF.

 

HOW TO RESET

While in the Leadwire Resistance Compensation Mode, 
you can reset the calibrated status to its ex-factory state by 
pressing both UP/DOWN switches at once.  Confirm that 
the LED turns off when you press these switches.

I/O RANGE MODIFICATION

Configurator Software is used to change input types and 
ranges.
For changing output ranges, set the Output Range Selector 
and then select by Configurator Software.

 

SIDE VIEW

Output Range Selector

V

I

I

V

I

Output Range Selector

OUTPUT RANGE

SW POSITION

V2Z1: -10 – +10 V / 0 – 20 mA DC

I

V2V2: -10 – +10 V / -10 – +10 V DC

V

STATUS INDICATOR LED

The unit is provided with a status indicator LED which 
blinks in different patterns indicating various status of its 
CPU.  When it is functioning within normal parameters, the  
LED blinks in a regular pattern of ON and OFF.
The following figure indicates typical patterns.

LED PATTERNS

STATUS

Normal

Programming Mode
Out. 1 Calibration Mode
Out. 2 Calibration Mode

Burnout

Leadwire Resistance

Compensation Mode

100 ms

ON

OFF

MAINTENANCE

Regular calibration procedure is explained below:

 

CALIBRATION

Warm up the unit for at least 10 minutes.  Apply 0%, 25%, 
50%, 75% and 100% input signal.  Check that the output 
signal for the respective input signal remains within accu-
racy described in the data sheet.  When the output is out of 
tolerance, recalibrate the unit according to the “CALIBRA-
TION” explained earlier.

LIGHTNING SURGE PROTECTION

M-System offers a series of lightning surge protector for 
protection against induced lightning surges.  Please contact 
M-System to choose appropriate models.

Содержание M8XR2

Страница 1: ...nfirm its operational range as indicated below 24V DC rating 24V 10 approx 50mA 70mA for current output GENERAL PRECAUTIONS Before you remove the unit or mount it turn off the power supply and input signal for safety ENVIRONMENT Indoor use When heavy dust or metal particles are present in the air install the unit inside proper housing with sufficient ventilation Do not install the unit where it is...

Страница 2: ... that all cables are correctly con nected according to the connection diagram 2 DIP switches Check the output range selector setting 3 Power input Check the power input voltage 4 Input Check voltage across the terminal A B with a sensitive voltmeter With 20 C or 68 F approx 110mV with Pt 100 approx 55mV with Pt 50 Ω If RTD wires are broken the output goes over 100 be low 0 with downscale due to bu...

Страница 3: ...figurator Jack DOWN Switch BOTTOM VIEW Zero Span Selector S Z OFF Z S Zero Span Selector MODE SW POSITION SPAN S OFF ZERO Z HOW TO CALIBRATE THE ZERO SPAN Use a signal source and measuring instruments of sufficient accuracy level Turn the power supply on and warm up for more than 10 minutes The output 1 and output 2 adjustments are linked except for 2 and 6 1 Apply 0 input 2 Select the Output 1 Ze...

Страница 4: ...gurator Software is used to change input types and ranges For changing output ranges set the Output Range Selector and then select by Configurator Software SIDE VIEW Output Range Selector V I I V I Output Range Selector OUTPUT RANGE SW POSITION V2Z1 10 10 V 0 20 mA DC I V2V2 10 10 V 10 10 V DC V STATUS INDICATOR LED The unit is provided with a status indicator LED which blinks in different pattern...