WIRING INSTRUCTIONS
■
SCREW TERMINAL
Torque
: 0.8 N·m
CHECKING
1 ) Terminal wiring: Check that all cables are correctly
connected according to the connection diagram.
2 ) Auxiliary power input voltage: Check voltage across
terminal 9 – 10 with a multimeter.
3 ) Input: Check that the input shows within 0 – 120% of
the rated voltage.
Be sure that the input waveform is sinusoidal to obtain
proper accuracy for model L2PA.
4 ) Output: Check that the load resistance meets the de-
scribed specifications.
ADJUSTMENT PROCEDURE
This unit is calibrated at the factory to meet the ordered
specifications, therefore you usually do not need any cali-
bration.
For matching the signal to a receiving instrument or in case
of regular calibration, adjust the output as explained in the
following.
■
HOW TO CALIBRATE THE OUTPUT SIGNAL
Use a signal source and measuring instruments of suffi-
cient accuracy level. Turn the auxiliary power supply on
and warm up for more than 10 minutes.
1 ) ZERO: Apply 0% input and adjust output to 0%.
2 ) SPAN: Apply 100% input and adjust output to 100%.
3 ) Check ZERO adjustment again with 0% input.
4 ) When ZERO value is changed, repeat the above proce-
dure 1) – 3).
5 ) Go through the above procedure for channels 2.
MAINTENANCE
Regular calibration procedure is explained below:
■
CALIBRATION
Warm up the unit for at least 10 minutes. Apply 0%, 25%,
50%, 75% and 100% input signal. Check that the output
signal for the respective input signal remains within accu-
racy described in the data sheet. When the output is out of
tolerance, recalibrate the unit according to the “ADJUST-
MENT PROCEDURE” explained earlier.
LIGHTNING SURGE PROTECTION
M-System offers a series of lightning surge protector for
protection against induced lightning surges. Please contact
M-System to choose appropriate models.