CHECKING
1) Terminal wiring: Check that all cables are correctly con-
nected according to the connection diagram.
2) Power input voltage: Check voltage across the terminal
7 – 8 with a multimeter.
3) Input: Check that the input signal is within 0 – 100% of
the full-scale.
4) Output: Check that the load resistance meets the de-
scribed specifications.
ADJUSTMENT PROCEDURE
This unit is calibrated at the factory to meet the ordered
specifications, therefore you usually do not need any cali-
bration.
For matching the signal to a receiving instrument or in case
of regular calibration, adjust the output as explained in the
following.
■
HOW TO CALIBRATE THE OUTPUT SIGNAL
Use a signal source and measuring instruments of sufficient
accuracy level. Turn the power supply on and warm up for
more than 10 minutes.
1) ZERO: Apply the input signals which should lead the
calculated result as 0%, and adjust output to 0%.
2) SPAM: Apply the input signals which should lead the
calculated result as 100%, and adjust output to 100%.
3) Check ZERO adjustment again.
4) When ZERO value is changed, repeat the above proce-
dure 1) – 3).
EQUATION
Output = K
1
×
Input 1 + K
2
×
Input 2
K
1
, K
2
: 0.10 to 2.00 (specified when ordering)
Output, Input 1, Input 2: 0 – 100%
Note: Be aware that a input signal less than 0 % is calculated
as a minus signal.
MAINTENANCE
Regular calibration procedure is explained below:
■
CALIBRATION
Warm up the unit for at least 10 minutes. Apply the input
signals which should lead the calculated results as 0%, 25%,
50%, 75% and 100%. Check that the output signal for the
respective input signal remains within accuracy described
in the data sheet. When the output is out of tolerance, rec-
alibrate the unit according to the “ADJUSTMENT PROCE-
DURE” explained earlier.
LIGHTNING SURGE PROTECTION
M-System offers a series of lightning surge protector for
protection against induced lightning surges. Please contact
M-System to choose appropriate models.
KADS
5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN
Phone: +81(6)6659-8201 Fax: +81(6)6659-8510 E-mail: [email protected]