LW MEASUREMENTS TREE PIZA+ 12 Скачать руководство пользователя страница 6

When to calibrate 
 

Note: Calibration is done in factory. Don't re-calibrate the Scale, unless it is not 

accurate. 

 

Calibration: 

 

A.  Turn on the power switch of the scale

Press and hold 

[ZERO]

 key then press 

[ON/OFF]

, the display 

will show 

“CAPU=”

, press twice 

[HOLD]

 key, the display will show 

CALU=

, press 

[UNIT]

 key to select 

calibration unit kg or lb, it will have a small triangle on the right to indicate kg or lb, press 

[HOLD]

 key 

to confirm the selection.  

 

B.  Press 

[HOLD]

 key again, the display will show 

CAL

, press 

[UNIT]

 key to 

set calibration weight. Press 

[UNIT]

 key to move the flash digit to right, then  

press 

[ZERO]

 key to increase the flash digit, press 

[HOLD]

 key to confirm  

the selection ( the calibration weight should be more than 

2/3

 of full  

capacity), it will show CAL.  

 

C.  Press 

[PRINT]

 , it will show CAL and AD value.  

 

D.  Press 

[UNIT]

 key, the display will show the flash calibration weight, place the shown weight on the 

platform, press 

[UNIT]

 key when the stable indicator brightens up, the display will show “

--------

”, then 

the AD value. Restart the scale and calibration has done. 

 

 

 

 

 

 

 

 

 

 

 

 

 

 