LW MEASUREMENTS LSS-400 Скачать руководство пользователя страница 4

TARE 

key

Tare Range (100% to full capacity)

ZERO 

key 

Set display to zero or 
Zero Range (4% to full capacity)

DISPLAY 

   

Low battery indicator             Weighing unit

          Negative value indicator
                                      

                                      
                                    
     
                                     

ZERO

 indicator       Hold indicator               

Indicates when the reading is 0

  

                                

2

HOLD FUNCTION

1)

Place the item on the scale platform. 

2)

Wait for the weights to be displayed.

3)

Press the [HOLD] key. The “

.

” hold indicator will turn-on.

4) Remove  the  item  from the   scale  platform.  The  item’s  weight  reading  will 

remain on the display. You can now also convert this to other measures (i.e. 
to Pounds or Kilograms) by pressing the [UNIT] key.

5) Quit   hold   function:   Once   an   item   is   weighed   and   the   hold   function   is 

enabled; you want to quit to hold function, Press the [HOLD] key again to 
cancel the “

.

” hold indicator and return the scale to weight mode. 

 

WEIGHING

Before weighing

Whenever possible, please allow the scale to warm up for several seconds 
after first turning the power on so that the scale will function properly and 
accurately.

Error messages

EEEE:    Overload

:     Low battery

Err-Z:   Exceed zero track range when turn on the power 
Err- S:    The scale is unstable when turn on the power
Err- C:    Forget to input calibration weight when calibrating or the AD value is 

too low.

Weighing procedures

1.

Press [ON/OFF] to turn on the scale.

When the power is turned on, all display segments appear for a few seconds 
and "0" will appear on the display.

2.

Select the weighing unit with [UNIT].

Press 

[UNIT]

 to select a weighing unit "kg", "lb".

Once the unit has been selected, the selected unit will be displayed next to the 
weight value.