Example:
If you want to convert the pupillary distance at 40cm optic distance when the
measured infinite far left or right monocular pupillary distance is 32.5mm and binocular
pupillary distance 65mm (as shown in Fig.13a), press Pupillary Distance Conversion
Key (TDIS) to choose the optic distance 40cm which needs to be converted, and
the converted values show that left or right monocular pupillary distance is 30.5mm
and binocular pupillary distance 61mm (as shown in Fig.13b).
Fig.13a
Fig.13b
Important Matters
The pupillary distance actually measured by the PM 200 is the pupillary distance
at the distance of
∞
far target, and pupillary distances at the distances of other targets
are obtained through calculation, only for reference.
4.
Maintenance
4.1
It has been precisely adjusted before delivery. Please do not dismantle it so as
to keep it precise.
4.2
It shall be stored and used in an indoor well-conditioned dry place.
4.3
As a high-tech product, the device shall be prevented from Vibration or impact.
4.4
Keep it clean and do not touch surface of its window glass. Fingerprint, dust
or stain should be cleaned with absorbent cotton dipped with mixed solution
of alcohol and ether.
4.5
Any corrosive chemical is forbidden for cleaning it.
4.6
In case of malfunction, do not dismantle it by yourself. Please contact the local
sales agent or the manufacturer for help.
4.7
Provided there is no display on the display window after startup, please check
the polarity of the battery to see if it is well placed and electrified.
16
Содержание PM 200