background image

At Operation 

- Never disassemble or touch the inside of the instrument. 

  This may result in an electric shock or instrument malfunction. 

-Never yank the power cord to disconnect from wall outlet but hold the plug while 

disconnecting. 

  This can weaken the metal core of the cord and may result in a short circuit or 

an electric shock. 

 

At Storage 

- Do not store the instrument in a place where it may get wet or where   

  poisonous gas or liquid is stored. 

- Avoid storing the instrument in an area with excessive heat, humidity, or dust. 

  Recommended ranges: 

     Temperature range within              -10

  to +70

 

     Relative humidity range within              30% to 85% 

     Atmospheric pressure range within     500 to 1060hPa 

 

 

At Installation 

- Do not install the instrument near water. 

If water gets into the internal structure, there is the possibility of electrical 

Shock or instrument malfunction. 

LUXVISION - LCP-2600

Содержание LCP-2600

Страница 1: ......

Страница 2: ...rior to use of this equipment Table of Contents Introduction 1 Package Contents 1 Classifications 2 Designated Symbols 3 Safety Instructions 4 Instrument Components 6 Installation 9 Focusing 10 Operating Instructions 11 Charts Description 13 LUXVISION LCP 2600 ...

Страница 3: ...Maintenance 16 Basic Trouble Shooting 19 Transportation 19 Technical Specifications 20 LUXVISION LCP 2600 ...

Страница 4: ...ow the steps in this manual in order to ensure safe operation optimum performance and a longer service life from your new instrument Please retain this manual for future reference LCP 2600 and accessories were carefully checked and packed prior to the shipment However please check condition and contents upon delivery In addition to LCP 2600 this shipment includes Package Contents 1 metal Screen 40...

Страница 5: ... shock ClassⅠ LCP 2600 is Classified into ClassⅠ instrument ClassⅠ instrument in which protection against electric shock does not rely on basic insulation only but which includes an additional safety precaution with the provision of the connection of the instrument to the protective earth conductor in fixed wiring of the installation in such a way that any accessible conductive parts cannot become...

Страница 6: ...o splash water on the instrument Degree of protection against flammability LCP 2600 is classified as an equipment not suitable to be used in a Potentially flammable anesthetic mixture with air or with oxygen or nitrous oxide Do not use near flammable materials Mode of operation Continuous operation Symbol Designated Symbols Description This symbol indicates a protective earth LUXVISION LCP 2600 ...

Страница 7: ...ructions are included in this User s Guide This symbol on the main switch indicates that the power is OFF This symbol on the main switch indicates that the power is ON Manufacturer Date of Manufacturer This symbol indicates a fuse Safety Instructions LUXVISION LCP 2600 ...

Страница 8: ...ck At Storage Do not store the instrument in a place where it may get wet or where poisonous gas or liquid is stored Avoid storing the instrument in an area with excessive heat humidity or dust Recommended ranges Temperature range within 10 to 70 Relative humidity range within 30 to 85 Atmospheric pressure range within 500 to 1060hPa At Installation Do not install the instrument near water If wate...

Страница 9: ... After Use If the instrument is not be used for a long time disconnect the power cord from the wall outlet Otherwise it may cause a tire Remove batteries from the battery case of the remote controller when the remote controller will not be used for a long time At Disposal Follow local governing ordinances and recycling plans regarding disposal or recycling of device components Check the specified ...

Страница 10: ......

Страница 11: ...1 Upper Case 2 Projection Lens 3 Window for remote detection 4 Table stand 5 Focusing Wheel 6 Pilot Lamp 7 Power Switch LUXVISION LCP 2600 ...

Страница 12: ...5 8 RS 232C Connector 9 Power Connector 10 Fuse Holders LUXVISION LCP 2600 ...

Страница 13: ...cal horizontal line single 12 Letters chart Letter Masking 13 Number chart 18 Children chart 14 Snellen chart 19 Vectograph chart 15 Program 20 Red Green Test 16 Lamp ON OFF 21 Direction 11 12 13 14 15 16 17 18 19 20 21 LUXVISION LCP 2600 ...

Страница 14: ...ine the refracting distance form the patient s eye to the screen and install the projection screen 3 Position the instrument at the same distance from the screen to the patient within the range between 2m and 7m To obtain longer refracting distances in small rooms a mirror or systems of mirrors can be used In this case a high quality front surface mirror is required LUXVISION LCP 2600 ...

Страница 15: ...lignment of the system The instrument is optimized when the projection screen is angled to direct light to the patient s head Place a mirror on the screen The light should project where the patient s head would be LUXVISION LCP 2600 ...

Страница 16: ...or on and using the remote control project the 0 05 K onto the screen 3 If using a wall mount it is best to locate a wall stud to support the weight of instrument 4 Adjust the position of the instrument forward or backward for sizing 5 Adjust for sharpness and clarity by turning the focusing wheel left or right as needed LUXVISION LCP 2600 ...

Страница 17: ...mote controller The charts have 31 selections 3 Isolating the visual acuity chart To use Vertical line masking Press To move the position right or left Press or To move the position up or down Press or To use Horizontal line masking Press To move the position right or left Press or To move the position up or down Press or To use Single letter masking Press To move the position right or left Press ...

Страница 18: ...ess one of the chart buttons to memorize 4 And then press the right button and heard buzzer sound 1 time to memorize the selected chart 5 Repeat the process to memorize the selection you want 6 Press the left button and heard buzzer sound 3 times to complete the installation of Program A To install Program B 1 Make sure press the button of LAMP to turn off the lamp 2 Keep pressing the right button...

Страница 19: ...d heard buzzer sound 3 times to complete the installation of Program B Programmed Operation 1 Turn on LCP 2600 press the button of RESET to select the program A or B 2 After operate the program change the chart by the button of PROGRAM Left move to previous chart Right move to next chart A TYPE CHARTS DESCRIPTION letters Duochrome Balance Cross Cylinder Dots LUXVISION LCP 2600 ...

Страница 20: ...Red Green Astigmatism Clock Dial Snellen Worth For Dot Phoria Fixation Phoria with Fixation Schober Coincidence Number Cross Grid Minute Stereo Children Binocular Balance LUXVISION LCP 2600 ...

Страница 21: ...PE letters Snellen Children Binocular Balance Duochrome Balance Cross Cylinder Dots Red Green Astigmatism Clock Dial Landolt Worth For Dot Phoria with Fixation Number Fixation Coincidence LUXVISION LCP 2600 ...

Страница 22: ...Schober Minute Stereo Cross Grid C TYPE letters Horizontal Line Red Green Cross Cylinder Dots Snellen Worth For Dot Astigmatism Clock Dial Fixation Phoria LUXVISION LCP 2600 ...

Страница 23: ... Balance Minute Stereo Children Vertical Line Risk of electric shock Always disconnect the power cord from the wall and the instrument prior to performing any of the following procedures MAINTENANCE Maintenance There is no periodic or routine user maintenance required LUXVISION LCP 2600 ...

Страница 24: ...me reason the instrument becomes soiled wipe it with the damp lint free cloth and mild detergent Then wipe off with a dry cloth to finish Disassembling the upper case 1 Turn off the power and pull out the power cord 2 Unfasten the screw in counterclockwise at the back side of case 3 Disassemble the Upper Case by lifting it 4 In case of reassembling place the Upper case on the main body and adjust ...

Страница 25: ...e heat 1 Unfasten the screw at the Lamp Cover 2 Disassemble the Lamp Cover by lifting it and pull out the Lamp 3 Place the new Lamp at the hole of Lamp Holder 4 Turn on the power switch 5 Place the white paper in front of the projection lens in order to project the filament image of lamp LUXVISION LCP 2600 ...

Страница 26: ... 1 Turn off the power switch and disconnect the power cord 2 Push and turn the fuse holder a quarter counterclockwise with your finger and remove it 3 Pull out the fuse from the fuse holder and check the condition 4 Replace the fuse with the provided spare or equivalent as specified below 5 Simply return it with the fuse holder to the socket Push and turn it a VOLTAGE FUSE 100V 240V 1A 250V T1AL L...

Страница 27: ...ource Circuit Breakers 2 Check the electrical connections Power Cord 3 Check the main fuse located on the rear of the unit If the projector lamp does not light 1 Check the main On Off Switch 2 Replace the lamp with the spare lamp The lamp may be HOT Do not touch the Lamp directly and allow the sufficient amount of time to cool down the heat If the projector turns on but does not function 1 Check t...

Страница 28: ...r electric bulb ex Halogen lamp it will improperly operate on the effect of electronic waves Technical Specifications Chart 31 Selections Chart selection speed Average 0 3 sec Mask 1 open 3 horizontal lines 3 vertical lines Filter Red Green Projection distance 2 0m 7 0m 3m is standard Projection magnification 30 x at 5m Projection size φ252mm and 330mm W x 225mm H at 5m Tilt angle Ball joint 10 de...

Страница 29: ...e controller 180g Accessories 1 dust cover 1 metal screen 400mm x 350mm 1 Pair of polarizing spectacles 2 spare fuses Optional accessories wall bracket floor stand red green glasses Interface 8 pin connector for simultaneous control of chart mask and R G filter by a motorized refractor s control unit LUXVISION LCP 2600 ...

Страница 30: ......

Отзывы: