LUXO LHH G2 Скачать руководство пользователя страница 6

Symbol explanations:

 

 

Do not throw away. Electronic waste. 

   

 

The product must be recycled

 

 

Self declaration symbol

 

 

Approved by Nemko

 

 

Attention, read instructions before use

Warning:

•  Avoid spilling liquid on the lamp head

•  Have minimum 400 Lux light, when reading the labels on the product

•  The power cable must only be replaced by a qualified electrical fitter or other similarly qualified professional.

•  The switch disengages the low voltage to the lamp only. To disconnect the mains supply, unplug the plug from the    

 

socket outlet.

•  To avoid risk of electric shock, this equipment must only be connected to a supply main with protective earth.

•  This ME EQUIPMENT could cause risk of reciprocal interference during specific investigations or treatments.

•  Position the product so that it’s easy to disconnect the device when the product is used as isolation means.

•  To terminate operation, turn the product off with the switch. For storage, unplug the device and fold the arm together.

•  Any repairs of the arm must be performed by Glamox or Glamox approved personnel. The arm contains high tensions  

 springs.

•  Do not modify this equipment without authorization of the manufacturer.

•  This product complies to IEC/EN 60601-1-2, electromagnetic compatibility for medical devices. Electromagnetic or    

 

other interference between this equipment and other devices should not occur. In the event of interference,

 

please move the equipments apart from each other.

•  The switch disengages the low voltage to the lamp only. To disconnect the mains supply, unplug the plug from the    

 

socket outlet.

•  When abnormal use Glamox declines all responsibilities regarding the product

•  Users of this form of lighting should be aware that the manufacturer of the product can only accept liability for the    

 

lighting when maintenance, repairs or changes are carried out by the manufacturer or by companies which the

 

manufacturer has approved to carry out such work, and when components that influence the safety of the lighting are  

 

replaced with original spare parts. The lighting is not intended for use in areas where there is a risk of explosion.

Disposal

Disposal the product at a special collection centre for recycling of electronic and waste.

Environmentally conditions

Transport and storage

 

Ambient temperature .................................................................... -20° C to 70° C

 

Moisture content .......................................................... 10 % to 100 % (keep dry)

 

Atmospheric pressure ............................................................ 500hPa til 1060hPa

Operate

 

Ambient temperature ....................................................................... 5° C to 40° C

 

Moisture content ................................................................................ 30 % to 75 %

 

Atmospheric pressure ............................................................ 700hPA Max 3000m 

6

LHH G2/LHH G2 CCT

Содержание LHH G2

Страница 1: ...Luxo LHH G2 LHH G2 CCT User manual...

Страница 2: ...e inzamelpunt gebracht worden in uw gemeente of regio Dansk Vores produkter er underlagt direktiv 2002 96 EC Waste Electrical and Electronic Equipment WEEE og skal efter endt levetid leveres til modta...

Страница 3: ...table bracket or floor stand Insert product and remove protective plastic sleeve or rubber band and carefully unfold the product Use only brackets supplied by Glamox or a Glamox representative Glamox...

Страница 4: ...unctionality if third party accessories are used Cleaning Blow and dry off any dust and dirt Using of different cleaning agents for cleaning this product Avoid use of alcohol acids or any kind of orga...

Страница 5: ...kage current Protection reliability Must not exceed 0 5 mA Power supply Resistance in earth circuit earth connection 0 1 Power consumption 100 240VAC 50 60 Hz Mode of operation LHH G2 max 60VA LHH G2...

Страница 6: ...of the manufacturer This product complies to IEC EN 60601 1 2 electromagnetic compatibility for medical devices Electromagnetic or other interference between this equipment and other devices should n...

Страница 7: ...onfirms with the standards listed Reference Date of issue Name EN 60601 1 2012 General requirements lighting EN 60601 1 6 2010 Collateral standard Usability EN 60601 2 41 2009 Particular requirements...

Страница 8: ...arranty Please visit our web site for contact information www glamox com Glamox March 2016 Item no 05 900 713 00 Doc no BRU 303 0112 Rev no 0 Manufactured by Glamox AS BU Luxo production Kirken r Indu...

Отзывы: