Summary

 

If velocity and displacement is of particular interest then use at frequencies 
above 10 Hz is desirable). During handling/mounting of the sensor for low 
frequency (below 10Hz), movement of the sensor/cable will affect  the meter 
readings and they should not be considered accurate. 

 Lutron state a 

Frequency Range of 10Hz to 1kHz

 and a calibration point of 

160Hz

 in the specification area of the manual. We recommend to have the 

anchor point of the cable on the fixture that is being measured (as per following 
page). This will allow the cable to move in the same motion as the sensor – and 
forces on the sensor and cable will now have a better chance of staying equal, 
this is more critical at lower frequencies 

 

 

 