pH

CONTROLLER/MONITOR

Model : PPH-2108

Your  purchase  of  this  pH  CONTROLLER/MONITOR
MONITOR marks a step forward for you into the field of
precision measurement. Although this pH controller is a
complex and delicate instrument, its durable structure
developed. Please read the following instructions carefully
and always keep this manual within easy reach.

       OPERATION MANUAL

Содержание PPH-2108

Страница 1: ...NITOR marks a step forward for you into the field of precision measurement Although this pH controller is a complex and delicate instrument its durable structure developed Please read the following in...

Страница 2: ...by a qualified technician only Power plug should apply the correct ACV power voltage Operating duty cycle should be adhered to Cleaning Only use the dry cloth to clean the plastic case Equipment prot...

Страница 3: ...cator 5 3 12 Wire terminals 5 3 13 RS232 terminal 5 3 14 BNC Input socket 5 3 15 Case holder 5 3 16 pH Electrode plug optional 5 3 17 pH Electrode optional 5 3 18 Temp Probe optional 5 3 19 Temp Wires...

Страница 4: ...ill be make action On Off when the reading value reach high limit or low limit value Temperature Offset value setting Hysteresis value setting for high and low alarm temp unit setting with default Lar...

Страница 5: ...surement ATC probe TP 07A 0 to 65 pH PH7 PH4 and PH10 3 points calibration Calibration ensure the best linearity and accuracy pH Optional Electrode Any PH electrode with BNC connector Temp Probe Optio...

Страница 6: ...VA AC 110V Consumption Approx 5 3 VA AC 220V Weight 384 g 0 84 LB Meter only Dimension DIN size 96 x 48 mm Depth 110 mm Accessories Instruction manual 1 PC Included Case holder with screw 2 PCs Option...

Страница 7: ...y 0 to 14 PH 0 01 PH 0 02 PH 2 d Temperature used optional Temp probe TP 07 A Measurement Range Resolution Accuracy 0 to 65 0 to 65 0 8 32 to 149 32 to 149 1 5 Specification tests under the environmen...

Страница 8: ...utton 3 6 Button 3 7 PH Temp Button 3 8 pH control relay indicator 3 9 Temp control relay indicator 3 10 pH indicator 3 11 Temp indicator 3 12 Wire terminals 3 13 RS232 terminal 3 14 BNC Input socket...

Страница 9: ...erminals 2 Connect the pH Control Relay output from T3 T4 Connect the Temp Control Relay output from T5 T6 4 2 pH calibration When the first time use the pH electrode or the pH electrode already be us...

Страница 10: ...dicator 3 11 Fig 1 will light the Display 3 1 Fig 1 will show the Temp value that sensing from the Temp probe ATC probe TP 07A Press the PH Temp Button 3 7 Fig 1 once again the Display will return the...

Страница 11: ...3 10 Fig 1 will light again The pH measurement value will be compensated by the Manual Temp value The default Manual Temp value is 25 The procedures to set the Manual Temp value refer to page 11 The m...

Страница 12: ...ght High Limit Value Setting 1 After set the Low Limit value press the Set Button 3 4 Fig 1 twice the Display will show HILt now the meter is ready for the pH High Limit value setting Press the PH Tem...

Страница 13: ...cedures tPty Temp unit setting tPSt Temp compensation value setting HySt Hysteresis setting tPoA Temp Offset setting PCPS Probe Calibration Period Setting Temperature unit setting 1 Press the Set Butt...

Страница 14: ...the Button 3 6 Fig 1 to adjust the desiring Temp compensation value If the meter already connect the Temp probe ATC probe TP 07A though already adjust the Temp compensation value the pH measurement va...

Страница 15: ...age 12 Fig 2 Fig 2 For example High limit value 100 Low limit value 20 Hysteresis value 5 a The control relay will On when measuring value up to 100 The control relay will Off again when measuring val...

Страница 16: ...the display reading will present warning message to inform the user should to execute the new calibration procedures 1 After finish the Temp offset value setting press the Set Button 3 4 Fig 1 twice t...

Страница 17: ...res should be executed by both buffer solution a pH 7 buffer solution b pH 4 buffer solution 3 It should be calibrated under the pH 7 buffer solution at first then calibrated under the pH 4 buffer sol...

Страница 18: ...onds the meter will entry the calibration value and finish the pH calibration procedures the Display will return to normal screen the PV indicator 3 2 Fig 1 will on the SV indicator 3 3 Fig 1 will off...

Страница 19: ...hing Use the pH 7 buffer solution it will show 7 00 Use the pH 4 buffer solution it will show 4 00 at the same time SV indicator 3 3 Fig 1 and the pH indicator 3 10 will light If the aim calibration v...

Страница 20: ...utilized for user s specific application A RS232 lead with the following connection will be required to link the instrument with the PC serial port Meter PC 3 5 mm jack plug 9W D Connector Center Pin...

Страница 21: ...r for Display 01 02 pH 05 D10 Polarity 0 Positive 1 Negative D9 Decimal Point DP position from right to the left 0 No DP 1 1 DP 2 2 DP 3 3 DP D8 to D1 Display reading D8 MSD D1 LSD For example If the...

Страница 22: ...times then if Display show oK means the quality of the electrode is oK if Display show Err means the electrode is already out of life and need to change the new electrode 8 SYSTEM RESET Power on the m...

Страница 23: ...9 THE ADDRESS OF AFTER SERVICE CENTER 20 0704 PPB2108...