background image

 

 

 

4)

 

To exit the Memory Record Function, press the “REC” Button for at least 2 seconds. The 

display will t

hen revert to the standard measurement screen and will no longer show “REC”

 

NOTE: To reset your Minimum or Maximum reading, whilst in the “REC MAX” or “REC MIN” 

screen press the hold button and this will delete the current Minimum or Maximum reading. 

5-4 Change Temperature Unit 

Press the “

°C/°

F” button (3

-4, Front Panel Description) continuously for a minimum of two 

seconds. Then release the button and the units will change from °C to °F or from °F to °C. 

6.

 

Data Logger 

The meter can save up to a maximum of 100 data points to its internal memory. These can then 
be reviewed on the screen. 

6-1 Saving Data 

1)

 

With the meter on press the “REC” Button, the display will show “REC” on the display.

 

2)

 

Press the “LOGGER” Button (3

-2, Front Panel Description) and the display will save one 

measuring value to memory. The display will then show you the position the data has been 
recorded to and how many memory points you have left. 

3)

 

To exit out of the data logging mode, press the “REC” button for at least two seconds. The 

display will then revert to the current 

reading and will no longer show “REC”.

 

6-2 Recalling Data 

1)

 

With the meter on press the “REC” Button, the display will show “REC” on the display.

 

2)

 

Press the “REC” Button continuously until the display shows you “TTL” and your relevant 

data point. 

3)

 

Use the      button or       to select which reading you wish to view. When you call the data 
the pH unit will flash and show you the result at the time the record was taken. 

4)

 

Press the “HOLD” button to exit out of the data recall menu.

 

6-3 Deleting Data 

1)

 

Ensure the meter is turned off. 

2)

 

Press the “REC” Button continuously and then press the “POWER” button momentarily. The 

display will show the following screen: 

 

 

 

 

 

Содержание PH-222

Страница 1: ...of precision measurement Although this meter is a complex and delicate instrument it offers durable structure that offer a long life span if correct calibration and storage techniques are implemented Please read the following instructions carefully and always keep this manual within easy reach ...

Страница 2: ...ompartment Cover 3 7 Protection cover 4 pH CALIBRATION PROCEDURE 4 1 Calibrating Consideration 4 2 Requiring Equipment for Calibration 4 3 Calibration procedures 5 MEASURING PROCEDURE 5 1 pH measurement 5 2 Data hold 5 3 Data record Max Min reading 5 4 Change Temp unit 6 DATA LOGGER 6 1 Save data 6 2 Call data 6 3 Delete data Empty the memory 7 REPLACEMENT OF BATTERY 8 REPLACEMENT OF pH ELECTRODE ...

Страница 3: ... Temperature sensor for Automatic Temperature Compensation ATC User replaceable pH electrode IP67 Rated Records Maximum and Minimum Values LCD with temperature and pH display along with bar graph indicator Automatic Power off functionality to save battery life Compact size and light weight 2 SPECIFICATIONS ...

Страница 4: ......

Страница 5: ...3 FRONT PANEL DESCRIPTION ...

Страница 6: ...edure pH 7 Calibration 1 Decant an appropriate amount of distilled water into a small beaker enough to cover the electrode bulb 2 Immerse the electrode and rinse it to remove any particulates from the sensor 3 Power on the instrument by pressing the POWER Button 3 2 Front Panel Description 4 Press the HOLD Button 3 3 Front Panel Description once and then press the button 3 4 Front Panel Descriptio...

Страница 7: ...0 Buffer solution PH 04A that came supplied Press the HOLD button once The display will show 4 00 flashing followed by END It will then return to the standard measurement screen and this concludes the pH 4 00 calibration process as shown below Calibration Clear Function If for some reason the calibration procedure is executed in the incorrect way the meter offers the ability to clear the meter of ...

Страница 8: ... you have completed your measurement rinse the electrode in distilled water and place it into the protection cap 3 7 Front Panel Description 5 2 Data Hold This meter has the ability to hold a particular measurement To do this during your measurement simply press the HOLD Button This will show a HOLD Symbol on the screen and will hold the current reading To release the hold measurement simply press...

Страница 9: ...the LOGGER Button 3 2 Front Panel Description and the display will save one measuring value to memory The display will then show you the position the data has been recorded to and how many memory points you have left 3 To exit out of the data logging mode press the REC button for at least two seconds The display will then revert to the current reading and will no longer show REC 6 2 Recalling Data...

Страница 10: ...scription Take out the depleted batteries and install 4 x AAA Alkaline Batteries ensuring correct polarity 3 After installing the batteries refit the battery cover 8 Replacement of pH Electrode 1 This meter features the ability to replace the electrode The electrode is the PE 12 and to remove the existing electrode unscrew the collar from the meter as per Figure 2 2 Position the new electrode with...

Отзывы: