background image

4. PRECAUTION & PREPARATIONS 
    FOR MEASUREMENTS

1. Remove the power from the circuit when making the

measurement. If any voltage is present in the testing
circuit, then an erroneous reading will result.

2. Ensure that the batteries (6 x 1.5 V AA battery) is

connected  correctly the right position into the battery
compartment.

3. Rotate the "Function/Range Switch" (3-2, Fig. 1) to the

right position before making measurement.

4. Slide the "Power Lock/Manual(off) Switch" (3-3, Fig. 1)

to the off position if the meter not be used.

5. MEASURING PROCEDURE

 

5-1  OHMS Measurement (Continuous Check)

*

Do not apply the external voltage
to the input terminals !

1. Connect the RED test plug into "ohm terminal"(3-5, 

Fig. 1)

2.

Connect the BLACK test plug into "COM terminal"(3-5, 
Fig. 1)

3. Rotate the "Function/Range Switch"(3-2, Fig. 1) to the

"200 OHM" position.

4. Connect test alligator clips into circuit under test.

5.

Push the "Test Button"(3-4, Fig. 1) for measurement.

Remark : 

A continuity beeper is equipped. If the measured resistance 

< approx. 50 ohm,  the beeper sounds will be generated.

5

Содержание DI-6300

Страница 1: ...a step forward for you into the field of precision measurement Although this INSULATION TESTER is a complex and delicate instrument its durable structure developed Please read the following instructi...

Страница 2: ...external voltage to input terminals at any time ACV ohms range do not apply the overload voltage to the input terminals Remove test leads before open the battery cover Cleaning Only use the dry cloth...

Страница 3: ...witch 4 3 4 Test Button 4 3 5 Input Terminal 4 3 6 Battery Cover Compartment 4 4 PRECAUTION PREPARATIONS FOR MEASUREMENTS 5 5 MEASURING PROCEDURE 5 5 1 OHMS Measurement Continuous Check 5 5 2 Mega OHM...

Страница 4: ...current 18mm 0 7 large LCD display Battery operating easily to carry with user LCD display allows clear read out even at bright place LSI circuit provides high reliability and durability Built in over...

Страница 5: ...ttery x 6 PCs Alkaline or heavy duty battery Power Approx 100 mA 1000M ohm 1000V range Consumption Approx 28 mA 200 M ohm 500V range Approx 28 mA 200 M ohm 250V range Approx 13 mA 200 M ohm 100V range...

Страница 6: ...00V 3 1d 0 1 M ohm 500 V 5 500 K ohm 1000 M ohm 1000V 3 1d 1 M ohm 1000 V 5 10 M ohm Input terminal Short circuit current is 2 55 mA OHMS Range Accuracy Resolution Open Circuit Voltage 200 ohm 1 1d 0...

Страница 7: ...2 FRONT PANEL DESCRIPTION Fig 1 3 1 Display 3 2 Function Range Switch 3 3 Power Lock Manual off Switch 3 4 Test Button 3 5 Input Terminal 3 6 Battery Cover Compartment 4...

Страница 8: ...de the Power Lock Manual off Switch 3 3 Fig 1 to the off position if the meter not be used 5 MEASURING PROCEDURE 5 1 OHMS Measurement Continuous Check Do not apply the external voltage to the input te...

Страница 9: ...nction Range Switch 3 2 Fig 1 to the 200M ohm 100V 200M ohm 250V 200M ohm 500V or 1000M ohm 1000V position according the user s requirement 4 Connect the test ALLIGATOR CLIPS into the circuit under te...

Страница 10: ...t should to slide the Power Lock Manual off Switch 3 3 Fig 1 to the POWER LOCK position Attention When the measurement the impedance resistance value less than 1 Mega ohm the period time of Power Lock...

Страница 11: ...shock 1 When the upper left corner of LCD display show BAT It is necessary to replace the battery However in spec measurement may still be made for several hours after LOW BATTERY INDICATOR appears be...

Страница 12: ...8 THE ADDRESS OF AFTER SERVICE CENTER 9 0805 DI6300...

Отзывы: