Acuity LS 

 

 Page 11 

Operating Manual  #OMA20200327 

 

 

Pause  Data  Logging 

–  Suspends  automatic  data  logging  and  instructs  the  Acuity  LS  device  to 

remain in low-power sleep mode.  

 

Download Data Log 

– Retrieves all stored data from the  Acuity LS device. Selecting this button 

will open a dialog window to save the downloaded data.  The default filename includes the Acuity 
LS device serial number, date, and time of the download.  The green indicator will blink while the 
data transfer is in progress. This may take several minutes. 

 

Delete  Data  Log 

–  Permanently  deletes  all  stored  data  on  the  Acuity  LS  device.  Selecting  this 

button will open a dialog window so that the user may confirm or cancel this request.  The  green 
indicator will blink until the deletion is complete. This may take several minutes.   

NOTE:  this  command  has  no  effect  on  other  stored  settings,  such  as  system  time,  logging 
interval, etc.   

CAUTION: this command will permanently erase all test data from the Acuity LS device. 

 

Switch  Acuity  LS  Device 

–  Allows  the  user  to  connect  a  different  Acuity  LS  device  without 

restarting the interface utility software. 

 

System Check 

– Instructs the Acuity LS device to immediately take a set of measurements and 

display  the  results  in  the  parameter  fields  below  in  the  main  screen.    This  command  is  used  to 
verifying  system  operation.  System  check  can  be  used  to  confirm  if  time  is  correct,  check 
remaining  battery  life,  and  verify  sensor  responses.        This  process  takes  about  20  seconds  to 
complete.      Points  collected  using  this  command  are  stored  in  the  Acuity  LS  memory,  but 
otherwise this command does not affect data logging. 

Clock Synchronization 

Each  Acuity  LS  device  has  an  internal  real-time  clock  that  maintains  device  time  using  Coordinated 
Universal Time (UTC+0).    

NOTE: User should make sure the connected computer time is accurate before synchronizing the 
Acuity LS clock. 

The user is prompted to synchronize the Acuity LS device time with the computer time whenever starting 
a data log or downloading data (Figure 9).  Use the Acuity LS interface program to synchronize the Acuity 
LS device clock with the connected computer clock.   

 

Figure 9.  Acuity LS device and computer time comparison window when starting the data logging or 

downloading data. 

 

 

Содержание Acuity LS

Страница 1: ...ACUITY CORROSION MANAGEMENT SOLUTIONS Acuity LSTM OPERATING MANUAL OMA20200327 sensinst lunainc com...

Страница 2: ...ment 6 System Restart 7 4 Software and Driver Installation 7 Acuity LS User Interface Software and USB Driver Installation 7 5 Software Operation 8 Establishing Communication 8 Operating the User Inte...

Страница 3: ...ble by the user Replacement Acuity LS Lid Sensor Panels LSP may be purchased from Luna 2 HARDWARE DEFINITION The Acuity LS system includes a multi sensor data acquisition device that saves data intern...

Страница 4: ...LSP and base interface 2 Remove all socket head cap screws using 9 64 hex wrench Take care to avoid sideways motion of the LSP as the screws are removed as this may damage the internal spring pin conn...

Страница 5: ...eck function in Section 5 Operating the User Interface will confirm the LSP is in good contact by displaying reasonable measurements for air temperature and relative humidity LSP installation is now c...

Страница 6: ...ttery cap Figure 4 3 Remove the metal battery contact tube and the battery For sensors mounted with battery cap facing up the battery is more difficult to extract In these situations it may be possibl...

Страница 7: ...tions for installing the software and drivers needed to communicate with Acuity LS Please ensure that your computer meets these basic system requirements Laptop or desktop computer with Windows 7 8 or...

Страница 8: ...e computer 1 Remove the protective cap Figure 1 item 4 and install the RS 485 cable Figure 1 item 5 The other end of the cable plugs into any USB 2 0 compliant serial port NOTE When finished communica...

Страница 9: ...port does not depend on which physical USB port is used Figure 6 Configuring the communications COM port 5 Select a COM port e g COM4 and click the OK button a If the correct COM port has been selecte...

Страница 10: ...Displays the sampling interval in minutes currently set for the device Acuity LS Device ID Displays the serial number of the connected Acuity LS device This number should match the label on the outsi...

Страница 11: ...connect a different Acuity LS device without restarting the interface utility software System Check Instructs the Acuity LS device to immediately take a set of measurements and display the results in...

Страница 12: ...the provided text boxes The Aircraft Asset window can be used to enter specific information and notes about the aircraft asset or structure being monitored The Acuity LS device window can be used to e...

Страница 13: ...to return to the main screen 8 Once data logging has started the System Status on main screen will indicate Logging Data and displays the selected Sampling Interval and Gold IDE Setup Download Data L...

Страница 14: ...ce first sample in data file C Air Temp C Temperature of the ambient air in degrees Celsius D RH Percent relative humidity of the ambient air E Surface Temp C Temperature of the structure in degrees C...

Страница 15: ...C 0 UNIX time may be readily converted to other time and date formats using the methods in Table 2 or this time converter https www epochconverter com Table 2 Converting UNIX time to other date and ti...

Страница 16: ...nits Sensor Excitation Surface Temperature Ts 40 85 C DC current Air Temperature Ta 40 85 C Relative Humidity RH 0 100 Conductance Low Freq GH 0 005 1 S 20 mVpp 10 Hz Conductance High Freq GL 5 10 000...

Страница 17: ...tem CD Compact Disk COM Communications FTDI Future Technology Devices International ID Identification IDE Interdigitated Electrode IPA Isopropyl Alcohol LSP Acuity LS Lid Sensor Panel PC Personal Comp...

Отзывы: