Service manual
Service Manual – 15 / 63
XE_ABS_MT_AN_200224
8.3.
Remove the screen from the base of the device
1/ Remove all the back and bottom cover of
the unit.
2/ Unplug all the cables that connect the
screen to the base of the device.
It concerns the following cables: CN24, CN25,
CN26, power supply cable, ground cable and
orange/green switch cable.
Unplug the cable:
CN24 from the USB board;
CN25 from the 20-5A ERM board;
CN26 from the 15-50 ERM board.
Then, take the cables out of the base.
3/ Unscrew the 2 screws at the back of the
screen.
4/ Take the screen out of the base.
Содержание Quantel ABSOLU
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 16: ...Service manual Service Manual 9 63 XE_ABS_MT_AN_200224 6 BLOCK DIAGRAM Superior part Inferior part...
Страница 63: ...Service manual Service Manual 56 63 XE_ABS_MT_AN_200224 8 When the procedure is finished click on Done...
Страница 66: ...Service manual Service Manual 59 63 XE_ABS_MT_AN_200224 15 4 Activate Axial Length...