Read this manual before operating your Mattress System.

Save this manual for future use.

The most current version of this manual can be found online at 

www.grahamfield.com.

SELECT SERIES 

LS200 ALTERNATING PRESSURE / 

LOW AIR LOSS MATTRESS SYSTEM

USER MANUAL

Содержание SELECT LS200

Страница 1: ...l before operating your Mattress System Save this manual for future use The most current version of this manual can be found online at www grahamfield com SELECT SERIES LS200 ALTERNATING PRESSURE LOW...

Страница 2: ...ATTRESS INSTALLATION 10 PUMP INSTALLATION 10 6 OPERATING INSTRUCTIONS 11 PATIENT PLACEMENT AND NURSING CARE 11 OPERATION 11 SKIN CARE 11 INCONTINENCE DRAINAGE 11 GENERAL OPERATION 11 STATIC FUNCTION 1...

Страница 3: ...g your Mattress System INTENDED USE The Lumex LS200 Alternating Pressure Low Air Loss Mattress System is designed to aid in the treatment and prevention of pressure ulcers while optimizing user comfor...

Страница 4: ...result in death or serious personal injury CAUTION Indicates a potential hazard situation or unsafe practice that if not avoided could result in minor or moderate personal injury s NOTICE Indicates a...

Страница 5: ...bed frame labeling or with the manufacturer for dimensions prior to mattress placement WARNING Side Rails Patient Restraints Whether and how to use side rails or restraints is a decision that should b...

Страница 6: ...G Power Cord Ensure power cord is kept free from all pinch points and moving parts and is not trapped under casters Improper handling of the power cord can cause damage to the cord which may possibly...

Страница 7: ...l on cell therapy whereby the cell is split in two where the bottom cells do not deflate if the upper cells are completely deflated in order to provide extra protection and zero pressure comfort for t...

Страница 8: ...ront panel Pump rear view MATTRESS FEATURES Individual air cushion design for maximum pressure distribution Vented air cushion provides true low air loss therapy Cell on cell design for zero pressure...

Страница 9: ...flow from the hose connection ports Place in plastic bag for storage Mattress Check the air manifold for kinks or breaks Replace if necessary Twist open the CPR plug at the head of the mattress and di...

Страница 10: ...s 4 Ensure mattress is properly positioned with no gaps between mattress and bed frame or side rails 5 Secure straps on bottom of mattress to the bed frame PUMP INSTALLATION POWER Switch Hanger Pump R...

Страница 11: ...cularly in areas where incontinence and drainage occur Ensure linens under patient are not wrinkled Early intervention may be essential to preventing serious skin breakdown Incontinence Drainage Use m...

Страница 12: ...on selected mode default mode is ALTERNATE 3 Allow time for full mattress inflation 4 After the mattress is fully inflated the caregiver can transfer the patient to the mattress Note the mattress can...

Страница 13: ...lacing your hand under the patient buttocks between cells and bottom of mattress The patient should have at least 1 1 2 in 4 cm of clearance between the buttocks and the bottom of the mattress ALTERNA...

Страница 14: ...rom bottoming out Info The low pressure alarm is automatically disengaged for 45 minutes when control unit is first powered on to ensure no alarms are raised during initial inflation CPR DEFLATION WAR...

Страница 15: ...ch Support at 1 770 368 4700 for further information When you believe a component or part is not functioning properly immediately contact GF Tech Support at 1 770 368 4700 as a potentially hazardous c...

Страница 16: ...m hypochlorite solution to dilution of 1000 ppm or any EPA approved hospital grade disinfectant do not use phenol base cleaning solution INSPECTION SYSTEM CHECK OUT Check each of the following before...

Страница 17: ...odification alteration or misuse may lead to serious injury and or product damage and will void all applicable warranties SYMPTOM POSSIBLE CAUSE SOLUTION Air is pumping out from the control unit but m...

Страница 18: ...Voltage Frequency AC 110 120V 60 Hz Fuse Rating T1AL 250V Maximum Current 0 1A Classification Electrical Class II Type BF Not AP or AGP type Temperature Operation 59 F 104 F 15 C 40 C Storage Transpo...

Страница 19: ...ts 4 Accessories or parts not provided by GF 5 Charges by anyone for adjustments repairs replacement parts installation or other work performed upon or in connection with such products which are not e...

Страница 20: ...LS200 INS LAB RevB17 20 11 EMC RELATED NOTIFICATION...

Страница 21: ...LS200 INS LAB RevB17 21...

Страница 22: ...LS200 INS LAB RevB17 22...

Страница 23: ...roduction 3 L Low pressure alarm 14 M Mattress cleaning 15 Mattress installation 10 N Notices 6 NOTICE statement significance 4 Nursing care 11 O Operating instructions 11 P Patient Placement 11 Pump...

Страница 24: ...e product information can be found on our website www grahamfield com 2017 GF Health Products Inc All Rights Reserved Graham Field and Lumex are trademarks of GF Health Products Inc GF Health Products...

Отзывы: