EMERGENCY RELEASE
In case of lift failure while using the Lumex LF500 Patient Lift, follow the procedure below to
safely lower the user. The Emergency Lowering Device is located at the top of the actuator
shaft. It is intended for use if the actuator fails to operate while a patient is suspended.
Pull down and
turn clockwise to lower
The device consists of a plastic collar ring that should be pulled down, then turned clockwise
continually until the patient has been lowered.
Contact your dealer immediately if standard troubleshooting techniques do not correct the
failure. Do not attempt to lift patients until all failure issues have been safely resolved.
Info: The emergency lowering device is intended for use during lift failure. This device
will allow lowering of patients only. Contact GF Tech Support at 1.770.368.4700 or your GF
authorized distributor immediately in case of failure.