LF1050-INS-LAB-RevH18 18
CLEANING
Regular cleaning is recommended every three months; clean the lift more often as needed. A soft
cloth, dampened with water and a mild NON-ABRASIVE detergent (household cleaner, soapy water,
or hospital grade disinfectant) is all that is needed to clean your patient lift. Automobile wax or
furniture polish will help maintain the finish over a long period of time.
s
NOTICE: DO NOT wash patient lift under water pressure or steam clean.
CARE OF SLINGS
Lumex slings are washable at 120°F or less. If there is a removable plastic insert in the head
support, remove insert before washing. Air dry.
s
NOTICE: DO NOT bleach slings!
MAINTENANCE SCHEDULE
Info: Only qualified persons should service and repair your Lumex LF1050 Patient Lift. Regular
maintenance of your patient lift is necessary to ensure continuing proper and safe operation.
Read and observe the following recommended maintenance schedule:
Item
Inspect
Inspect / Adjust
initially
monthly
every 3
months
Boom & spreader bar
Check connections between 1) boom and spreader bar and 2) boom and mast
for improper connection, looseness, or wear
X
X
Inspect boom for bending or deflection
X
X
Mast
Inspect mast for bending or deflection
X
X
Inspect steering bar for damage or loose hardware
X
X
Base and foot pedal
Ensure all hardware is tight
X
X
Ensure casters and axle bolts are tight
X
X
Inspect rubber parts of casters for deflection
X
X
Apply grease to caster ball bearings if needed
X
Actuator and control box
Ensure control box is firmly affixed to the mast
X
X
Ensure pins that connect actuator to boom and mast are firmly affixed
X
X
Ensure cable that connects actuator and control box is firmly affixed
X
X
Ensure battery indicator has at least 2 green LEDs illuminated before use
X
X
Sling and sling hardware
Check sling for wear; discard if worn
X
X
Ensure sling hardware is in good condition before each use
X
X