background image

LF1030-INS-LAB-RevB10  

12

Charging the battery with lift-mounted battery charger: 

With control box, battery charger, and 

battery assembled, insert the battery charger plug into a standard 110V receptacle. A green light 
signifying power and a yellow light signifying that the battery is charging will illuminate on the bat-
tery charger. When the battery is fully charged, the yellow light will shut off. The lift is now ready 
for use.

Note: The patient lift will not operate during lift-mounted battery charging.

Charging the battery with wall-mounted battery charger: 

Remove the battery from its bracket on 

the patient lift by placing your hand on top of the battery, curling your fingers behind the battery, 
and pulling up on the metal handle. Pull the battery up and toward you to remove. Place the bat-
tery on top of the wall-mounted battery charger. A green light signifying power and a yellow light 
signifying that the battery is charging will illuminate on the battery charger. When the battery is 
fully charged, the yellow light will shut off. When battery charging is complete, place the battery 
back on the patient lift. Ensure that the tabs on the battery are fully engaged with the bracket. The 
lift is now ready for use.

Emergency Procedures (LF1040 only)

Emergency stop: The control box is equipped with a red Emergency Stop button. This button is to 
be used in the event the hand control ceases to function correctly. To engage, hit the button firmly 
with the palm of your hand. The button will lock in the depressed position, stopping all motor 
functions. To disengage, turn the button clockwise, and it will return to its normal position.

Emergency lowering: 

The motor is equipped with an 

Emergency Release

, which is located on top of the motor 

as shown in the picture at right. To engage, pull up on the 
release, which will disengage the gearing in the cylinder, 
and lower the patient. The emergency release is adjust-
able by turning the screw adjacent to the pull stem. Turn 
the screw clockwise to increase and counter-clockwise to 
decrease the sensitivity of the release mechanism.

  WARNING: Use caution when using the emergency 

release. Pull back slowly on the release, which will 
gradually lower the patient. Pulling back too quick-
ly could cause the patient to be lowered too rap-
idly, which could result in serious injury. Do not, un-
der any circumstances, use the emergency release 
for lowering the patient when the hand control and 
battery are operating correctly. The emergency re-
lease should not be used in lieu of the hand control 
and battery.

motor

adjustment

   screw

emergency

release

emergency release

Содержание LF1030

Страница 1: ...s Inc Read this manual before operating your Patient Lift Save this manual for future use The most current version of this manual can be found online at www grahamfield com LF1030 HYDRAULIC PATIENT LI...

Страница 2: ...CAUTIONS 4 WARNINGS 4 3 FEATURES 6 4 HANDLING PROCEDURES 7 SHIPPING AND TRANSPORTATION INSTRUCTIONS 7 5 ASSEMBLY INSTRUCTIONS 8 COMPONENTS 8 ASSEMBLY 9 6 OPERATING INSTRUCTIONS 13 LIFTING THE PATIENT...

Страница 3: ...with proper supervision and with a capable individ ual acting as patient It is extremely important to know how a patient will feel in the patient lift We suggest that the operator try all the various...

Страница 4: ...O NOT exceed the maximum weight capacity WARNING GF Health Products Inc assumes no responsibility for any damage or injury caused by improper assembly or use of this product WARNING Check all parts fo...

Страница 5: ...electronics and actuator contain no serviceable components If service is required the unit must be returned to the factory for repair DO NOT attempt to open the electronics or actuator or obtain local...

Страница 6: ...ent Lift only Mast Push handle 6 point spreader bar suspension Hydraulic pump LF1030 Hydraulic Patient Lift only See above left for LF1040 Electric Patient Lift components Boom Locking casters Base ad...

Страница 7: ...s If damage is evident notify the car rier or your authorized Lumex Dealer 2 Remove all loose packing from the carton 3 Carefully remove all the components from the carton Note Unless the patient lift...

Страница 8: ...COMPONENTS LF1030 hydraulic patient lift components LF1040 1 3 2 4 7 5 8 6 Item Qty Description 1 1 Mast Boom Pump assembly LF1030 1 Mast Boom Actuator assembly LF1040 Note Actuator Charger Pendant a...

Страница 9: ...ted in the same direction as the base legs insert the mast into the sleeve 4 Insert the locking pin through the holes of the sleeve and mast and attach the locking cable to the end of the pin 5 Screw...

Страница 10: ...stall the base adjustment handle in the socket located behind the mast sleeve on the base With base handle to the far right position insert and tighten base attachment bolt to ensure that the handle i...

Страница 11: ...l plug Immediately to the right of the hand control connection is the motor connection Insert the hand control plug firmly in its receptacle this plug has nothing to align Charging the battery LF1040...

Страница 12: ...is now ready for use Emergency Procedures LF1040 only Emergency stop The control box is equipped with a red Emergency Stop button This button is to be used in the event the hand control ceases to func...

Страница 13: ...lock or block the patient lift casters when lifting The casters must be free to roll so that the patient lift can stabilize as the patient is lifted from the bed WARNING To ensure stability while lif...

Страница 14: ...t gently lean the patient forward Place the folded sling behind the patient s back and push it down until it touches the seat of the chair 3 Use the loops to pull the leg sections of the sling forward...

Страница 15: ...d attendant pull sling to smooth it and center patient on sling 5 Pull the leg loops forward and under the thighs 6 Cross the loops one through the other 7 With base of lift under bed SLOWLY release p...

Страница 16: ...nd bolts are tight The lift moves freely with caster brakes off Rear caster brakes operate properly Base adjustment mechanism operates easily Spreader bar clips are secured tightly The hydraulic pump...

Страница 17: ...placement parts CARE OF SLINGS Lumex slings are washable at 150 F and autoclavable at 185 F for 30 minutes The head support s plastic insert should be removed before washing The slings can be dried by...

Страница 18: ...ar right Use a hex bolt washers and hex nut to secure pump to mast Use two 9 16 wrenches to tighten 2 Attach the top of the hydraulic pump to the upper boom bracket as shown at right Use a hex bolt wa...

Страница 19: ...wear X Mast Ensure that mast is fully engaged and locked into mast sleeve X X Inspect mast for bends or deflections X Inspect all mast hardware for wear X X Boom Inspect boom for bends or deflections...

Страница 20: ...tery Battery charger defective Replace battery charger Battery defective Replace battery Hand control defective Replace hand control Actuator does not run but relay is clicking Actuator plug not prope...

Страница 21: ...PAIR REPLACEMENT CREDIT AND OR REFUND This warranty does not include any labor charges incurred in replacement part s installation or any associated freight or shipping charges to the manufacturer For...

Страница 22: ...and you may also have other legal rights which vary from state to state province to province Some states provinces do not allow the exclusion or limitation of incidental or consequential damage or lim...

Страница 23: ...the patient from a bed 13 Lifting the patient from the floor 14 Lowering the patient onto a bed 13 Lowering the patient to the floor 14 Lubrication 16 M Maintenance schedule 19 N Note significance 3 O...

Страница 24: ...USA Company Headquarters GF Health Products Inc 2935 Northeast Parkway Atlanta Georgia 30360 USA tel 800 347 5678 770 368 4700 fax 800 726 0601 678 291 3232 Made in Taiwan www grahamfield com...

Отзывы: