background image

LF1030-INS-LAB-RevB10 

© 2010 GF Health Products, Inc.

Read this manual before operating your Patient Lift.

Save this manual for future use.

The most current version of this manual can be found online at 

www.grahamfield.com

LF1030 HYDRAULIC PATIENT LIFT

LF1040 ELECTRIC PATIENT LIFT

USER MANUAL

Содержание LF1030

Страница 1: ...s Inc Read this manual before operating your Patient Lift Save this manual for future use The most current version of this manual can be found online at www grahamfield com LF1030 HYDRAULIC PATIENT LI...

Страница 2: ...CAUTIONS 4 WARNINGS 4 3 FEATURES 6 4 HANDLING PROCEDURES 7 SHIPPING AND TRANSPORTATION INSTRUCTIONS 7 5 ASSEMBLY INSTRUCTIONS 8 COMPONENTS 8 ASSEMBLY 9 6 OPERATING INSTRUCTIONS 13 LIFTING THE PATIENT...

Страница 3: ...with proper supervision and with a capable individ ual acting as patient It is extremely important to know how a patient will feel in the patient lift We suggest that the operator try all the various...

Страница 4: ...O NOT exceed the maximum weight capacity WARNING GF Health Products Inc assumes no responsibility for any damage or injury caused by improper assembly or use of this product WARNING Check all parts fo...

Страница 5: ...electronics and actuator contain no serviceable components If service is required the unit must be returned to the factory for repair DO NOT attempt to open the electronics or actuator or obtain local...

Страница 6: ...ent Lift only Mast Push handle 6 point spreader bar suspension Hydraulic pump LF1030 Hydraulic Patient Lift only See above left for LF1040 Electric Patient Lift components Boom Locking casters Base ad...

Страница 7: ...s If damage is evident notify the car rier or your authorized Lumex Dealer 2 Remove all loose packing from the carton 3 Carefully remove all the components from the carton Note Unless the patient lift...

Страница 8: ...COMPONENTS LF1030 hydraulic patient lift components LF1040 1 3 2 4 7 5 8 6 Item Qty Description 1 1 Mast Boom Pump assembly LF1030 1 Mast Boom Actuator assembly LF1040 Note Actuator Charger Pendant a...

Страница 9: ...ted in the same direction as the base legs insert the mast into the sleeve 4 Insert the locking pin through the holes of the sleeve and mast and attach the locking cable to the end of the pin 5 Screw...

Страница 10: ...stall the base adjustment handle in the socket located behind the mast sleeve on the base With base handle to the far right position insert and tighten base attachment bolt to ensure that the handle i...

Страница 11: ...l plug Immediately to the right of the hand control connection is the motor connection Insert the hand control plug firmly in its receptacle this plug has nothing to align Charging the battery LF1040...

Страница 12: ...is now ready for use Emergency Procedures LF1040 only Emergency stop The control box is equipped with a red Emergency Stop button This button is to be used in the event the hand control ceases to func...

Страница 13: ...lock or block the patient lift casters when lifting The casters must be free to roll so that the patient lift can stabilize as the patient is lifted from the bed WARNING To ensure stability while lif...

Страница 14: ...t gently lean the patient forward Place the folded sling behind the patient s back and push it down until it touches the seat of the chair 3 Use the loops to pull the leg sections of the sling forward...

Страница 15: ...d attendant pull sling to smooth it and center patient on sling 5 Pull the leg loops forward and under the thighs 6 Cross the loops one through the other 7 With base of lift under bed SLOWLY release p...

Страница 16: ...nd bolts are tight The lift moves freely with caster brakes off Rear caster brakes operate properly Base adjustment mechanism operates easily Spreader bar clips are secured tightly The hydraulic pump...

Страница 17: ...placement parts CARE OF SLINGS Lumex slings are washable at 150 F and autoclavable at 185 F for 30 minutes The head support s plastic insert should be removed before washing The slings can be dried by...

Страница 18: ...ar right Use a hex bolt washers and hex nut to secure pump to mast Use two 9 16 wrenches to tighten 2 Attach the top of the hydraulic pump to the upper boom bracket as shown at right Use a hex bolt wa...

Страница 19: ...wear X Mast Ensure that mast is fully engaged and locked into mast sleeve X X Inspect mast for bends or deflections X Inspect all mast hardware for wear X X Boom Inspect boom for bends or deflections...

Страница 20: ...tery Battery charger defective Replace battery charger Battery defective Replace battery Hand control defective Replace hand control Actuator does not run but relay is clicking Actuator plug not prope...

Страница 21: ...PAIR REPLACEMENT CREDIT AND OR REFUND This warranty does not include any labor charges incurred in replacement part s installation or any associated freight or shipping charges to the manufacturer For...

Страница 22: ...and you may also have other legal rights which vary from state to state province to province Some states provinces do not allow the exclusion or limitation of incidental or consequential damage or lim...

Страница 23: ...the patient from a bed 13 Lifting the patient from the floor 14 Lowering the patient onto a bed 13 Lowering the patient to the floor 14 Lubrication 16 M Maintenance schedule 19 N Note significance 3 O...

Страница 24: ...USA Company Headquarters GF Health Products Inc 2935 Northeast Parkway Atlanta Georgia 30360 USA tel 800 347 5678 770 368 4700 fax 800 726 0601 678 291 3232 Made in Taiwan www grahamfield com...

Отзывы: