Страница 1: ...SPECTRA III L Light Engine Instruction Manual Lumencor Inc Document Number 57 10017 www lumencor com ...

Страница 2: ...y energy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio communications Safety Certifications TUV SUD America CB Certification IEC 61010 1 2010 TUV SUD America NRTLus Certification UL 61010 1 2012 05 TUV SUD America cNRTL Certification CAN CSA C22 2 No 61010 1 2012 TUV SUD America EN Certification EN 61010 1 2010 Underwriters Laborato...

Страница 3: ...utions and Warnings 3 Installation and Operating Instructions 4 Light Output Characteristics 5 Operational Specifications 6 Routine Maintenance and Trouble Shooting 7 Customer Support 8 Warranty SPECTRA III L Light Engine Instruction Manual 3 ...

Страница 4: ...ut over time The light sources within the SPECTRA III L light engine are controlled by an on board microprocessor operating Lumencor firmware accessed via one of two serial interfaces USB RS 232 or TCP The user can enable or disable each source independently by serial commands as well as change the intensity of each source independently The SPECTRA III L light engine can either be controlled by th...

Страница 5: ...nt déclarations qui identifient des situations ou des pratiques susceptibles d entraîner des blessures corporelles Caution Statements identify conditions or practices that could result in damage to your equipment Attention déclarations qui identifient des situations ou des pratiques susceptibles d endommager le matériel Safety Items Mesures de sécurité Warning ONLY use the power supply provided by...

Страница 6: ...er la lumière sans l extrémité de sortie du guide de lumière dirigée en toute sécurité dans un chemin optique fermé NE PAS pointer la sortie de lumière directement sur un matériau susceptible d être inflammable ou susceptible de brûler Cela comprend tous les tissus les plastiques les tissus le papier et les liquides animaux ou végétaux RISK GROUP 3 Warning Possibly hazardous optical radiation emit...

Страница 7: ... rayonnement laser visible ou dispersé visible et invisible Laser Aperture Warning OR Warning Avoid exposure laser radiation is emitted from this aperture Do not turn on light engine without first connecting a light guide to the output aperture The distal end of this light guide must be coupled into an enclosed optical path prior to operation Each operational control and laser aperture that can be...

Страница 8: ...lled liquids may damage your light engine Attention NE PAS placer de liquide sur la source lumineuse Les liquides renversés peuvent endommager la source lumineuse Caution DO NOT drop the light engine It contains glass optical components that could be damaged or misaligned by the shock produced by a drop onto a hard surface Attention NE PAS laisser tomber la source lumineuse Elle contient des compo...

Страница 9: ...er Figure 2 6 Quickstart Guide instruction document 85 10049 SPECTRA III L Light Engine Instruction Manual 9 AC Power Cords Region Part Number North America 29 10002 Europe 29 10005 United Kingdom 29 10004 Israel 29 10008 Australia New Zealand 29 10024 Figure 2 Control key remote interlock jumper and external gate jumper These items are packed in a plastic bag inside the shipping box Their functio...

Страница 10: ...F and are locked in this state until the internal temperature is below 50 C and or the fan restarts The current reading of the onboard temperature sensor is displayed on the front panel status display Figure 3 and in the Control GUI Section 3 3 3 3 3 Operation 3 3 1 Controls and Interlocks The Master Power Switch button on the front panel Figure 3 turns the electrical power to the unit on or off A...

Страница 11: ...the master power button 3 7 Wait 30 45 seconds for the initiation sequence onboard microprocessor boot up to complete Do not press any buttons or insert any plugs during this time 8 When the initiation sequence completes LUMENCOR will flash on front display panel and then be replaced by a display showing the current light engine IP address the internal temperature and the fan status At the same ti...

Страница 12: ...Pv4 6 Click Properties button 7 Use the following IP addresses IP Address 192 168 201 201 Subnet Mask 255 255 255 0 Default gate way and DNS Server are OK to leave blank 8 Connect the RJ45 cable supplied with the light engine between the LAN port on the light engine and an ethernet port on the computer 9 Type the Light Engine IP address Figure 3 into any web browser address bar to access the resid...

Страница 13: ...e Web GUI control interface as described in Section 3 3 3 3 Go to the SETTINGS page in the Web GUI Make sure that the USB port configuration is set to LEGACY mode and USB 5V is set to ENABLED 4 The pod must be set in SPECTRA III L control mode The light engine control mode setting is shown in green letters at the bottom of the pod display screen If the pod is not in SPECTRA III L control mode chan...

Страница 14: ... their center wavelength and maximum output power Not all sources may be turned on simultaneously To prevent exceeding the capacity of the DC power supply power consumption is tracked by the onboard computer If a set limit see Governor Figure 5 is exceeded either by increasing intensity settings for sources that are already on or by turning on additional sources commands will be rejected To clear ...

Страница 15: ...n accordance with the specifications listed on the certificate of conformance please contact Lumencor Technical Support for assistance as directed in Section 7 SPECTRA III L Light Engine Instruction Manual 15 Specification Detail Temperature Operating 32 to 86 F 0 to 30 C Non operating 4 to 158 F 20 to 70 C Humidity Operating and non operating 0 to 80 relative humidity non condensing Altitude Oper...

Страница 16: ...check all of these 1 Control key must be inserted in front panel and turned to on position 2 Liquid light guide must be inserted in light output receptacle 3 Remote interlock jumper must be inserted in rear panel 4 External gate jumper must be inserted in front panel Figure 3 Unusually weak fluorescence signals across all detection channels Weak fluorescence in all detection channels DAPI FITC TRI...

Страница 17: ...ing conditions consistent with the specifications given in Section 5 and observing all the Precautions and Warnings notified in Section 2 This warranty does not extend to light engines that have been subject to misuse accident tampering or improper installation Accessories including but not limited to liquid light guides optical fibers collimators cables and control consoles are not covered by the...