background image

GET IN TOUCH

Lumencor, Inc.

14940 NW Greenbrier Parkway, Beaverton, OR 97006 USA  •   T 503.213.4269  •  www.lumencor.com

©

2016 Lumencor, Inc.  • Effective Date: 3/21/2016 • 54-10036

Light Engine Control Pod

Set Up and Operating Instructions

FIGURE E: 

Color levels display

Operation (Local Mode)

1. Press 

the 

LEFT

 button to select the desired color channel. Successive presses will cycle through the available color channels.

2. Press 

the 

RIGHT

 button to turn the selected light source 

ON

. Adjust the output intensity using the rotary dial. Press the 

RIGHT

 button again to turn the

selected light source 

OFF

.

3.  Press and hold the 

LEFT

 button to access the color levels display and view digital output intensity settings for each color channel [1]. Channels that are

currently 

ON

 are highlighted with a white background. Press the 

RIGHT

 button to return to the main display screen.

Operation (Pass-through Mode)

1. Serial commands sent from a host computer to the USB B port of the control pod will automatically switch the pod from local to pass-through mode,

indicated by the message “PC pass through mode active” shown on the pod display screen.

2.  Local command mode is disabled as long as pass-through mode remains active.

3.  To quit pass-through mode and return to local command mode, press the 

RIGHT

 button on the pod twice [2].

Notes

[1] The current intensity settings are internally stored.  When the pod is powered down, the settings are retained and will be restored at the next restart.
[2] Changes to channel 

ON/OFF

 status and intensity levels made on the host computer during pass-through operation will be automatically replicated in

the pod settings.