GET IN TOUCH
Lumencor, Inc.
14940 NW Greenbrier Parkway, Beaverton, OR 97006 USA • T 503.213.4269 • www.lumencor.com
©
2016 Lumencor, Inc. • Effective Date: 3/21/2016 • 54-10036
Light Engine Control Pod
Set Up and Operating Instructions
FIGURE E:
Color levels display
Operation (Local Mode)
1. Press
the
LEFT
button to select the desired color channel. Successive presses will cycle through the available color channels.
2. Press
the
RIGHT
button to turn the selected light source
ON
. Adjust the output intensity using the rotary dial. Press the
RIGHT
button again to turn the
selected light source
OFF
.
3. Press and hold the
LEFT
button to access the color levels display and view digital output intensity settings for each color channel [1]. Channels that are
currently
ON
are highlighted with a white background. Press the
RIGHT
button to return to the main display screen.
Operation (Pass-through Mode)
1. Serial commands sent from a host computer to the USB B port of the control pod will automatically switch the pod from local to pass-through mode,
indicated by the message “PC pass through mode active” shown on the pod display screen.
2. Local command mode is disabled as long as pass-through mode remains active.
3. To quit pass-through mode and return to local command mode, press the
RIGHT
button on the pod twice [2].
Notes
[1] The current intensity settings are internally stored. When the pod is powered down, the settings are retained and will be restored at the next restart.
[2] Changes to channel
ON/OFF
status and intensity levels made on the host computer during pass-through operation will be automatically replicated in
the pod settings.
FIGURE F:
Pass through mode display