4.2 How to check/clean X-Cite® 120XL/Q metal vents

1.  When unit is OFF and lamp is COOL, unplug the power from the X-Cite® unit and turn the unit upside 

down.  Visually inspect the metal gratings for lint, dust and other debris.   

2.  Debris can be removed from the gratings using a small vacuum cleaner with a crevice tool, or blown  

out with a can of compressed air.  If using compressed air, direct the air flow ACROSS the vents, not 

perpendicularly (i.e. into the X-Cite® unit).   

 

4.3  How to check/clean X-Cite® 120 (original design) nylon foam filters  

1.  Visually inspect the perforations on the side of the unit, under the lamp access door.  You should be  

able to see a grey mesh filter on the inside of the unit.  This filter requires cleaning or replacing if:  

Dust/lint/debris has accumulated 

• 

Dust/lint/debris is blocking the view of the grey mesh filter 

• 

The filter looks as though it is crumbling 

• 

2.  It is possible to clean newer filters, but it is generally better to replace them, as they tend to deteriorate 

over time (one reason they stopped being used in X-Cite® units in 2006).  Contact Lumen Dynamics 

Technical Support for additional information on cleaning/replacing filters.    

 

4.4 What happens if filters are not clean?

A complete filter blockage or airflow obstruction generally results in an automatic lamp shutoff within 10-15 

minutes of the unit being powered on due to lack of adequate cooling.   

A partially clogged filter may allow enough airflow for continued operation, but not for optimal cooling.  In this 

case, the result is usually a dramatic reduction in lamp life, e.g. 400 hours instead of 2500+ hours, it may also cause 

bubbles to form in the light guide.  

 

Distributeur : 

Distributed by : 

Fast Drying Systems (FDS) SA 

  Tel:   +41 (0) 22 788 84 88 

 +33 (0) 4 58 00 10 07 

[email protected]

 

www.fluorescence.ch

 

Содержание X-Cite 120 Series

Страница 1: ...signal is low to begin with or is excited by a low power portion of the X Cite lamp s output spectrum 2 Liquid Light Guides 2 1 Proper handling and installation of liquid light guides Liquid light gui...

Страница 2: ...e when the other end is 4 connected to the X Cite unit if the unit is turned on in this condition the cap will overheat melt and or permanently discolour the quartz end of the light guide Do not expos...

Страница 3: ...bles can sometimes recover Disconnect the light guide from the X Cite unit and leave the light guide undisturbed on a shelf for 1 2 weeks For this to be effective it is important to catch the bubble w...

Страница 4: ...ife If the lower power regions of the output spectra are being used it may be desirable to replace lamps more frequently to maximize signal levels Elapsed Time Hours 0 200 400 600 800 1000 1200 1400 1...

Страница 5: ...amp will prevent this 3 Avoid unnecessary movement and jarring of the lamp especially when it is in 4 operation or hot When changing a lamp 5 Handle the lamp only by the ceramic areas a Never touch th...

Страница 6: ...air flow between the X Cite and walls or other equipment 2 Do not remove the rubber feet on the X Cite unit or otherwise reduce block the space between the bottom of the unit and bench top This may c...

Страница 7: ...res cleaning or replacing if Dust lint debris has accumulated Dust lint debris is blocking the view of the grey mesh filter The filter looks as though it is crumbling 2 It is possible to clean newer f...

Отзывы: