Series IN 500 Operating Manual
Controls and Connections 19
3.6 Exchange of sensor head
Note:
Whether a sensor head can be used with a different electronic box is indicated
by the identifier "S1" on the sensor cable. See section 3.5.1 for more information.
The sensor head is exchangeable if a
sensor head with other optical data or a
longer cable is required. Each sensor
head cable is marked with a label
containing a sensor code which has to be
set in the pyrometer.
The sensor head cable is available in
lengths of 3 or 15 m which can be cut to
any required length. The sensor code
label should be affixed to the remaining
part of the cable.
The cable has to be prepared as shown in
the illustration. The shield must be
pushed above the washer and O-rings,
then the cable can be pushed into the converter and fixed with the clamp screw (series 510) or
soldered to the pins (series 520). (For cable connection, see schematic in section
3.2,
Connection cable
.
The sensor code at the cable has to be set in the pyrometer as follows:
Setting via software: Set these codes S1 and S2
in
“Pyrometer parameters“
under the
button
and confirm
them with “
OK
”. Now the new sensor head is
matching to the converter and the pyrometer is
ready for measuring.
Additionally in the IN 510 or IN 520, the setting of the sensor head codes can be done via
internal push buttons: Remove the cover of the converter and press the buttons and
simultaneously. While pressing these two buttons, push the “
PAR
” button too. The
display shows code S1. With the arrow buttons and
, the new code S1 has to be set
and confirmed with the “
PAR
” button. Now code S2 appears on the display and has to
be set and confirmed in the same way.
A setting of these codes with the HT 6000 or digital display DA 6000 is not possible.