IMPAC pyrometers IS 210 · IGA 210

 

 

 

 

3.1 

Wiring example if using a digital display with integrated power supply: 

 

 

 

3.2 

Example for wiring using an external power supply: 

 

 

 

 

Note:

  

Additional analyzing instruments, e.g. controllers, recorders, etc  
can be connected in series as shown in drawing above.

 

 
 

3.3 

Connection via USB adapter onto a PC (for pyrometer parametrizing) 

 

Controller 

°C

Recorder 

Digital display

230 V 

~

4 ... 20 mA 

white

brown 

Schirm (schwarz)

24 V DC

Power supply

+ green
- yellow 

Power supply 
for LED targeting light 

White: 

+24 V 

Pin 1 

Brown: 

0 V 

Pin 2 

Black: 

Shield 

 

Green: 

LED + 

Pin 5 

Yellow: 

LED - 

Pin 4 

1

2

3

4

5

plug insert 
(plug side) 

Controller 

°C

Recorder 

Digital display  
with built-in  
2-wire supply 

230 V 

~

white

brown

Shield (black)

4 ... 20 mA 

White: 

+24 V  Pin 1 

Brown: 

0 V 

Pin 2 

Black: 

Shield   

Green: 

LED +  Pin 5 

Yellow: 

LED - 

Pin 4 

+ green

- yellow 

Power supply 
for LED targeting light 

1

2

3

4

5

plug insert 
(plug side) 

Power supply  
as displayed 
above 

USB adapter

USB port PC

Содержание IGA 210

Страница 1: ...IS 210 IGA 210 IMPAC Pyrometer Proven Quality Operation Manual Betriebsanleitung...

Страница 2: ...a USB adapter onto a PC 6 3 4 Connection to the portable parametrizing device HT 6000 7 4 Mechanical Installation accessories 7 5 Optics 7 6 Alignment with the LED targeting light 8 7 Parameter Settin...

Страница 3: ...s that arise from the use of any exam ples or processes mentioned in this manual or in case the content of this document should be incomplete or incorrect LumaSense Technologies reserves the right to...

Страница 4: ...peratures 75 C the output signal is set to approx 3 9 mA Load max 700 at 24 V power supply 400 18 V 1000 30 V Service interface for parameter programming with portable setup device HT 6000 or via USB...

Страница 5: ...dditional devices operating with this mains voltage e g transformers Touching mains voltage can be fatal An incorrect connection and mounting can cause serious health or material damages Only qualifie...

Страница 6: ...corder Digital display 230 V 4 20 mA white brown Schirm schwarz 24 V DC Power supply green yellow Power supply for LED targeting light White 24 V Pin 1 Brown 0 V Pin 2 Black Shield Green LED Pin 5 Yel...

Страница 7: ...least as big as the spot size of the pyrometer in the meas uring distance The following drawings show the pyrometer s spot sizes M in different measuring distances a The spot size M in a distance of a...

Страница 8: ...ch are different from the nominal distance of the selected optics may be calculated using the following formula D D M a a M 2 2 D D M a a M 1 1 6 Alignment with the LED targeting light For exact measu...

Страница 9: ...r the two spectral ranges of the in struments are listed in the emissivity table below The tolerance of the emissivity values for each material is mainly dependent on the surface conditions Rough surf...

Страница 10: ...until a new hot object appears in the measuring beam The temperature which has to be recognized as hot is defined by the low limit of the adjusted sub range The stored maximum value will be deleted wh...

Страница 11: ...surement curves Subsequent analysis graphically or in table form for print out or export Spot size calculator to calculate spot sizes in any distances 9 Transport packaging storage With faulty shippin...

Страница 12: ...suring distance see 5 Emissivity set too high Set lower correct emissivity corresponding to the material see 7 2 Lens contaminated Clean lens carefully see 10 2 Temperature indication too high Emissiv...

Страница 13: ...60 Connection cable 5 m 3 821 770 Connection cable 10 m 3 821 780 Connection cable 15 m 3 821 790 Connection cable 20 m 3 821 800 Connection cable 25 m 3 821 810 Connection cable 30 m 3 826 500 Portab...

Отзывы: