Operating Manual CHM 15k
R13 / 07-2019
Commissioning and Decommissioning
23
Command
Description
Response
(shortened)
get<SPACE>16:L<CR><LF>
Output of extended data telegram
see 8.3.4
set<SPACE>16:RNO=14<CR><LF>
Sets the device RS485 address
from 16 to 14
set 16:RNO=14
set<SPACE>16:Baud=4<CR><LF>
Sets baud rate to 19,200
set 16:Baud=4
set<SPACE>16:dt(s)=15<CR><LF>
Sets logging time interval to 15 s
set 16:dt(s)=15
get<SPACE>16:Lifetime(h)<CR><LF>
Readout of the laser operating
hour counter
get 16:Lifetime(h)
Table 3
Commands for functional testing (examples).
After completing the simple functional test on the CHM 15k:
•
Continue to operate in polling mode,
or
•
Reset to automatic transmission mode
set<SPACE><RS485Number>:Transfermode=1<CR><LF>
Comment: This command leads back to the automatic transmission mode of the standard telegram "1".
Baud rate with raw data transmission
The baud rate settings must be observed, especially in RS485 bus mode. If raw data
transmission is required, each telegram can be 12 kB in size. To reduce the
transmission time between two 15 s telegrams the baud rate shall be set to at least
19,200 baud.
7.2 Starting up with the LAN connection
In addition, or as an alternative to the RS485 connection, a LAN connection (Ethernet) can also be used.
Requirement: A connected LAN cable (see 6.2 Electrical installation) or LAN over DSL connection with
a DSL transmitter and receiver modem in between.
Configuration
: 3 separate IP addresses are available for simultaneous communication:
1. A preconfigured fixed service address to the device
➔
192.168.100.101, subnet 255.255.255.0
2. DHCP server assignment (requires a DHCP server)
3. User a gateway,
see section 8.7 Communication via Ethernet web interface for the configuration with a LAN
/ WAN connection to the device and 8.1 and 8.2, if these values are configured with the
RS485 interface.
The service address (1) cannot be changed by the user. It is always available and can be used as a direct
connection between a laptop and the CHM 15k Nimbus.
One of the 3 IP addresses can be entered into a web browser (Figure 15) for communication with the
device. Figure 20 shows the "Config Network" tab in the Internet browser Firefox. Changing the user IP
address (3) requires superuser authorization in the device tab.
The superuser password is:
15k-Nimbus
The superuser password can be changed if necessary, see Figure 22.