Communication RS485 & Ethernet
56
Revision R09, November 2016
8.7
Communication via Ethernet - Web-Interface
The communication with the CHM 15k Nimbus using the Ethernet connection is
a fast, secure and system independent way. Inside the instrument runs an
Apache web server. It enables a communication and configuration platform via
web interface to handle firmware updates, quick views of measuring results or
downloading full day NetCDF raw data files.
Figure 14 shows the start screen after a successful connection to the instrument
(start-up procedure in section 7.2).
In general the web interface does include the following right management:
l
End-user can check the state of the instrument.
l
Super user can additionally download NetCDF files, configure the instru-
ment, update firmware, download user manual and further configuration
files.
l
Service user can set the instrument serial number, download the service
manual and upload configuration files.
By entering a password in the “Administration” form field, downloadable and
configurable elements are enabled, e.g. the IP network configuration (Fig. 17).
The status info on device panel and the process warning panel are displaying
warnings and errors updated once a minute. Listed codes in the status info are
corresponding to service codes in table 16. The process warning panel (figure
22) contains more details for service people.
In super user or service mode the device panel contains buttons to switch off or
restart the system.
Figure 15 shows the NetCDF files panel. NetCDF files can be selected and down-
loaded, which is visible for all users.
Fig. 14: Web interface device menu in standard user mode. The field is displayed
as the first, and shows status information. A login as superuser or service
user is available.