Model 42-49 Neutron Detector
Technical Manual
Section 5
Ludlum Measurements, Inc.
Page 5-2
December 2017
For example, an assumed operating voltage is 1100 volts, based upon the
flattest part of the plateau. The count rate at that voltage is 2380 counts
per minute (cpm), and the neutron field dose-equivalent rate is 20 rem/hr.
The sensitivity is calculated as:
hr
mrem
cpm
hr
mrem
cpm
y
Sensitivit
/
/
119
/
20
2380
This value should be approximately 120 cpm/mrem/hr.
Gamma Rejection Check
With the Model 42-49 connected to the Model 2200, adjust the Model
2200 HV to the assumed operating voltage determined above.
Remove the Model 42-49 detector from the moderator and place in a
10 R/hr
137
Cs gamma radiation field.
Take a one-minute count. If more than 10 counts are observed for the
one-minute period, decrease the operating voltage until the count rate
drops below 10 cpm; however, ensure that the HV remains in the
plateau region determined above.
Ensure that sensitivity for the selected operating voltage is
approximately 120 cpm/mrem/hr.
Return the Model 42-49 to the moderator.