Model 52-1 Family
Technical Manual
Section 4
Ludlum Measurements, Inc. Page 4-4 July 2018
Alarm Time:
adjustable from 1 to 255 seconds. This parameter controls
the duration of the RADIATION ALARM and is indicated by a steady
tone from audio.
Incomplete Time:
adjustable from 1 to 255 seconds. This parameter
controls the duration of INCOMPLETE, which is indicated by a low-
pitched, beeping sound.
RDA Size:
This number, in nCi, indicates the size of the source used
during the automatic RDA calculation. (The Model 52-1 does not utilize
the automatic RDA feature, and the RDA size is set to 500.)
Low Background (BKGND) Alarm:
This parameter sets the number of
counts per 0.2-second interval that the background should not fall below
in normal operation. If background falls below this level, an
INSTRUMENT FAILURE alarm will occur. This alarm indicates if a
detector has failed or is not connected. This alarm will also indicate if the
ambient background has decreased significantly.
High Background (BKGND) Alarm:
This alarm is the opposite of the
LOW BACKGROUND alarm. This alarm will indicate that background
radiation has increased or that a detector has failed.
Sigma:
This parameter sets the number of standard deviations above
background that it takes to set off an alarm. A standard deviation is
assumed to be equal to the square root of the background count. As a
simple example, a 3-sigma alarm of a detector with a background count
rate of 100 counts is calculated as 3.0 times the square root of 100, or 30.
In this simple example, a sudden count of 130 counts would cause an
alarm.
Detector Activity:
This parameter notifies the user whether or not the
detector is active, indicated with a “Y” for “yes” or an “N” for “no.”
Individual Alarms:
An alarm can be set individually for each detector.
This parameter indicates whether one is set and informs the user with a
“Y” for “yes” or an “N” for “no.”
Sum Alarms:
This parameter informs the user whether a sum alarm, a
contribution from multiple detectors, is set, indicated with a “Y” for
“yes” or an “N” for “no.”
# of Samples:
In walk-through mode, this number determines how many
0.2-second samples to group for each alarm calculation.
BKG Update:
This number, adjustable from 1 to 9, adjusts how fast the
background is updated while the portal is unoccupied. The BKG
UPDATE number, multiplied by 18 seconds, gives the response time of
the internal background accumulation.
Содержание 52-1
Страница 3: ......
Страница 5: ......
Страница 6: ......
Страница 42: ......
Страница 43: ......
Страница 44: ......
Страница 45: ......
Страница 46: ......
Страница 47: ......
Страница 48: ......
Страница 49: ......
Страница 50: ......
Страница 51: ......
Страница 52: ......
Страница 53: ......
Страница 54: ......
Страница 55: ......
Страница 56: ......
Страница 57: ......
Страница 58: ......
Страница 59: ......
Страница 60: ......
Страница 61: ......
Страница 62: ......
Страница 63: ......
Страница 64: ......
Страница 65: ......
Страница 66: ......
Страница 67: ......
Страница 68: ......
Страница 69: ......
Страница 70: ......
Страница 71: ......
Страница 72: ......
Страница 73: ......
Страница 74: ......
Страница 75: ......
Страница 76: ......
Страница 77: ......
Страница 78: ......
Страница 79: ......
Страница 80: ......
Страница 81: ......
Страница 82: ......
Страница 83: ......
Страница 84: ......
Страница 85: ......
Страница 86: ......
Страница 87: ......
Страница 88: ......
Страница 89: ......
Страница 90: ......
Страница 91: ......
Страница 92: ......
Страница 93: ......
Страница 94: ......
Страница 95: ......
Страница 96: ......
Страница 97: ......
Страница 98: ......
Страница 99: ......