Model 4525 Generation IV Series                          Installation Manual

   

Section 2

 

 

Ludlum Measurements, Inc. 

Page 2-3 

December 2020 

 

If a computer is purchased with the system, it will be inside the crate. 

 

Any other optional items will be packed in the crate with the detectors. 

 

One crate is approximately 681 kg (1500 lb), and the other is approximately 454 kg 
(1000 lb). 

 

C

RATE 

C

ONTENTS OF THE 

M

ODEL 

4525-10000

 WITHOUT STANDS

  

The system is shipped in two crates.  

The first crate contains: Detector 1 and Detector 2 and peripherals. 

 

Two  large  (approximately  116.8  x  116.8  x  22.9  cm  {46  in  x  46  in  x  9  in.})  white 
detector boxes labeled Detector 1 and Detector 2 

 

One 30.5 m (100 ft) roll of cable for the detectors (Belden 8774 or equivalent) 

 

One 76.2 m (250 ft)  roll of CAT-5E cable for Ethernet connection (Belden 1501A 
or equivalent)  

 

One remote display with 30.5 m (100 ft) cable (Belden 9946 or equivalent) 

 

Four infrared sensor components with conduit  

 

One hardware mounting kit containing the following:  

 

16 qty. ½-inch x 2 ¼-inch long hex bolts  

 

16 qty. ½-inch hex nuts 

 

16 qty. ½-inch lock washers 

 

32 qty. ½-inch flat washers 

 

If a computer is purchased with the system, it will be inside the crate. 

 

Any other optional items will be packed in the crate with the detectors. 

 

One crate is approximately 681 kg (1500 lb), and the other is approximately 454 kg 
(1000 lb). 

 

The second crate contains: Detector 3 and Detector 4. 

 

Two large (approximately (meters)m {46 in x 46 in x 9 in.}) white detector boxes 
labeled Detector 3 and 4. 

 

Содержание 4525 GENERATION IV Series

Страница 1: ...LUM MODEL 4525 GENERATION IV SERIES RADIATION PORTAL MONITOR MODELS 4525 5000 4525 7500 4525 10000 4525 12500 4525 15000 INSTALLATION MANUAL December 2020 Serial Number GM5179 and Succeeding Serial Numbers ...

Страница 2: ...LUM MODEL 4525 GENERATION IV SERIES RADIATION PORTAL MONITOR MODELS 4525 5000 4525 7500 4525 10000 4525 12500 4525 15000 INSTALLATION MANUAL December 2020 Serial Number GM5179 and Succeeding Serial Numbers ...

Страница 3: ......

Страница 4: ... fitness which extend beyond the description of the face thereof If the product does not perform as warranted herein the purchaser s sole remedy shall be repair recalibration or replacement at the discretion of Ludlum Measurements In no event will Ludlum Measurements be liable for damages lost revenue lost wages or any other incidental or consequential damages arising from the purchase use or inab...

Страница 5: ...conduit s between each detector and the control box with 24 inches 61 cm over pull on each end Systems with more than 2 detectors run cable and conduit between the primary and secondary detectors Mount Infrareds as shown in site layout drawings provided in the installation manual and in loca tions specified on each IR assembly determined by system purchased Install Cat 5e Ethernet cable 250 foot 7...

Страница 6: ... over pull 1 each If applicable Photos of the horn strobe unit and or camera and or traffic light and or Wireless P2P system both antenna mounted with unobstructed signal path and connecting cables on each end mounted with conduit s and entering the control box 2 4 each Photo s showing the mounted computer printer and network equipment and cables with Cat5e cable connecting exterior detector if us...

Страница 7: ...ation 2 8 Typical Stand Setup 2 10 Setting Up The Stands 2 10 Conduit Requirements 2 15 Cable Requirements 2 15 Wiring 2 16 Detector Wiring 2 17 Remote Display 2 18 Cat 5E LAN Cabling 2 18 Supervisor Computer 2 19 Miscellaneous Requirements 2 19 Software and Network Installation 3 Software Installation 3 1 Network Installation 3 2 Software License Agreement 3 8 Drawings and Diagrams 4 ...

Страница 8: ...he detectors are normally mounted to the side of the vehicle and also sometimes above the vehicle The detectors are centered around the typical load your situation or location might require the detectors to be higher or lower than the dimensions shown in the back of this manual Each system comes with a wall mount computer display printer and universal power supply UPS This computer logs data and d...

Страница 9: ...tion 4 Typical Stand Setup 5 Setting Up The Stands 6 Conduit Requirements 7 Cable Requirements 8 Remote Display 9 CAT 5E LAN Cabling 10 Supervisor Computer optional 11 Monitor Computer optional 12 Miscellaneous Requirements Note Drawings of the model numbers listed above are located in the back of this manual Before building and installing the stands please review the drawings of the Model 4525 th...

Страница 10: ...ide the crate Any other optional items will be packed in the crate with the detectors The total shipping weight of the crate is approximately 681 kg 1500 lb CRATE CONTENTS OF THE MODEL 4525 7500 WITHOUT STANDS The system is shipped in two crates The first crate contains Detector 1 and Detector 2 Two large approximately 116 8 x 116 8 x 22 9 cm 46 in x 46 in x 9 in white detector boxes labeled Detec...

Страница 11: ...e 30 5 m 100 ft roll of cable for the detectors Belden 8774 or equivalent One 76 2 m 250 ft roll of CAT 5E cable for Ethernet connection Belden 1501A or equivalent One remote display with 30 5 m 100 ft cable Belden 9946 or equivalent Four infrared sensor components with conduit One hardware mounting kit containing the following 16 qty inch x 2 inch long hex bolts 16 qty inch hex nuts 16 qty inch l...

Страница 12: ...ion Belden 1501A or equivalent One remote display with 30 5 m 100 ft cable Belden 9946 or equivalent Four infrared sensor components with conduit One hardware mounting kit containing the following 20 qty inch x 2 inch long hex bolts 20 qty inch hex nuts 20 qty inch lock washers 40 qty inch flat washers If a computer is purchased with the system it will be inside the crate Any other optional items ...

Страница 13: ...en 1501A or equivalent One remote display with 30 5 m 100 ft cable Belden 9946 or equivalent Four infrared sensor components with conduit One hardware mounting kit containing the following 24 qty inch x 2 inch long hex bolts 24 qty inch hex nuts 24 qty inch lock washers 48 qty inch flat washers If a computer is purchased with the system it will be inside the crate Any other optional items will be ...

Страница 14: ...he system comes on two pallets The smaller pallet contains two large yellow I beam frame assemblies with white detectors attached For a Model 4525 7500 the pallet holds one detector per frame assembly For the Model 4525 12500 and 15000 the pallet holds 2 detectors per frame assembly On a Model 4525 7500 the pallet contains Detector 1 and 2 For the Model 4525 12500 and 15000 the pallet contains Det...

Страница 15: ...t buffer zone can be maintained around the detectors Typically the stands are mounted about 3 m 10 ft before the weigh scale It is important that vehicles stay out of this buffer zone except when they are moving slowly between the detectors Placing a stop sign at the entrance to this buffer zone is recommended as well as painting a line on the pavement See Drawing 511 x 928P Failure to correctly l...

Страница 16: ...ember 2020 Stand Orientation Detector 1 is connected to the office supervisor computer and mains power Detector 1 should go on the side with the four conduits coming out of the pad The other detectors are only connected to Detector 1 See figures below for layout MODEL 4525 5000 MODEL 4525 7500 ...

Страница 17: ...Model4525GenerationIVSeries InstallationManual Section2 Ludlum Measurements Inc Page 2 9 December 2020 MODEL 4525 10000 MODEL 4525 12500 MODEL 4525 15000 ...

Страница 18: ...t may be necessary to install additional steel or lead behind the detector enclosure to reduce background Setting up the Stands ANCHOR BOLT PATTERN Due to the close tolerances of the anchor bolt holes the anchor bolts must be placed according to the dimensions specified on the Anchor Bolt Template drawing 511 x 997 LIFTING There are 1 x 4 inch rectangular cutouts in the frame pieces that can be us...

Страница 19: ...placed in the concrete Double nut the anchor bolt to the template before the concrete cures to ensure proper anchor bolt alignment LEVELING 1 Once the concrete is dry and all the bolts are straight the mounting points will need to be leveled Shims may be used but the following instructions assume the use of leveling nuts 2 Run a set of leveling nuts on the bolts as low as allowable one nut per bol...

Страница 20: ... alternate method the detector can be mounted to the back plate and then the detector and back plate can be lifted together and set into place with a crane or other suitable device using straps that are properly rated The weight of one detector and back plate is approximately 680 kg 1500 lb Systems purchased with stands Model 4525 5000 and 4525 7500 side stands The weight of one side of the stand ...

Страница 21: ...ame and carefully lift it into position Attach the frame with the appropriate bolts 4525 15000 The weight of one overhead frame with two detectors and two back plates attached is approximately 1588 kg 3500 lb Position properly rated lifting straps around and under the frame and carefully lift it into position Attach the frame with the appropriate bolts TORQUE SPECS FOR INCH BOLTS NUTS NOT SUPPLIED...

Страница 22: ...he detectors to the stands are inch either Grade 3 zinc plated steel or stainless steel or better Typical tightening torque is 69 ft lb SQUARING The opposing front outside corners of the stands should be within a half inch of each other on the diagonal reading Be sure to take the readings at the base of each set of stands see figures below ...

Страница 23: ...om out of the concrete up to the enclosures should be LIQUID TIGHT METALLIC Cable Requirements AC POWER Incoming power from the breaker box for the control box enclosure will be 14 gauge wire for the hot and neutral connections and a 6 gauge wire for the ground connection OVER PULL FOR TERMINATION All wires will be pulled leaving a 0 61 m 24 in tail past the end of the conduit Caution Caution must...

Страница 24: ...hite 13 2 Black from Brown Black 12 1 Brown from Brown Black 11 10 Pin MTA Det 1 2 or 3 10 Black from Black Yellow 10 9 Yellow from Black Yellow 9 8 Black from Blue Black 8 7 Blue from Blue Black 7 6 Black from Green Black 6 5 Green from Green Black 5 4 Black from White Black 4 3 White from White Black 3 2 Black from Red Black 2 1 Red from Red Black 1 Retain Orange Black and Red Green pairs for fu...

Страница 25: ...nd Black from Blue Black and Blue wire pairs are cut off and not used Interconnect to IR Sensor Wiring This is for units with 2 sets of IRs For units with 4 sets duplicate this wiring for the other sensors Master Primary 1 Receiver Incoming Side Wire Color MTA Label B Blue Grey 3 Cut Black Wire Incoming Side White 2 Brown 1 Receiver Outgoing Side Blue 3 Cut Black Wire Outgoing Side A White 2 Brown...

Страница 26: ...ling will be run in conduit to the office or booth where the remote display is mounted Once inside conduit is not required but could be requested by the Location Supervisor Cables outside of conduit will be secured every 0 30 m 1 ft to prevent entanglement and possible damage to the cable The cable will be run into the remote box leaving a 0 61 cm 2 ft tail past the inside of the box CAT 5E LAN Ca...

Страница 27: ...amp power source of 120 240 Vac LMI recommends dedicated 10 20 amp breakers for each lane and clearly labeling each breaker AC power should be delivered to the Master Detector and to the computer in accordance with code requirements which supersede these instructions Such requirements for example may require a disconnect device in clear sight of the hard wired control box Power entering the Master...

Страница 28: ...ecommended 64 GB hard drive for storing the database and the images SVGA monitor capable of a resolution of 640 x 480 or greater Sound card with speakers for audible voice alarms Note Before installing any new software please read the Ludlum Measurements software license agreement at the end of this section and un install all previous versions The Supervisor will install automatically when the CD ...

Страница 29: ...ed to connect all Model 4525s workstations and cameras optional Each computer workstation must have a 100 MB Ethernet network interface card NIC installed IP ADDRESSES Each device on a network has a unique address This number is called the Internet Protocol IP Address This number can either be dynamically or statically assigned A static IP address is assigned to the device and will not change A dy...

Страница 30: ... 168 200 229 MODEL 4525 NIC CONFIGURATION If the Model 4525s have never been configured use the IP Setup Tool exe to search for and display the current IP address configuration Select each Model 4525 found and set the IP address according to the table above The network mask is always 255 255 255 0 The Gateway and DNS can be left at their default values The baud rate should be set to 57600 Click Se...

Страница 31: ...codes used in wiring RJ45 eight position modular plugs The American National Standard Institute Telephone Industry Association Electronics Industry Association ANSI TIA EIA wiring standards allow both of these color codes The only difference is that the orange and green pairs are interchanged Because it provides backward compatibility for both one pair and two pair of Universal Service Order codes...

Страница 32: ...IVSeries InstallationManual Section3 Ludlum Measurements Inc Page 3 5 December 2020 T 568A STRAIGHT THROUGH ETHERNET CABLE The T 568A Standard should be used in all new installations T 568B STRAIGHT THROUGH ETHERNET CABLE ...

Страница 33: ...B standard Ethernet Cable Instructions Note Do not deform bend stretch staple run parallel with power cables or run Ethernet cables near noise inducing components The total length of wire segments between a device and a hub switch or between two devices cannot exceed 100 meters 328 ft for 100Base TX and 300 m 984 ft for 10Base T Looking at the connector with the pins at the top and facing you pin ...

Страница 34: ...contact with the pins 6 Align Pin 1 of the cable with Pin 1 of the connector and insert the cable into the connector Push firmly so the wires go all the way to the top of the connector and the jacket goes into the connector by about 0 48 cm 0 19 in Again check to make sure the colors are in the right order No more than 1 3 cm 0 5 in of the Ethernet cable should be untwisted otherwise it will be su...

Страница 35: ...Model4525GenerationIVSeries InstallationManual Section3 Ludlum Measurements Inc Page 3 8 December 2020 ...

Страница 36: ...Model4525GenerationIVSeries InstallationManual Section3 Ludlum Measurements Inc Page 3 9 December 2020 ...

Страница 37: ...Model4525GenerationIVSeries InstallationManual Section3 Ludlum Measurements Inc Page 3 10 December 2020 ...

Страница 38: ...928I Model 4525 12500 DETECTOR ELEVATION Drawing 511 928J Model 4525 12500 SENSTIVE AREA Drawing 511 x 928K Model 4525 15000 TYPICAL INSTALLATION Drawing 511 x 928L Model 4525 15000 DETECTOR ELEVATION Drawing 511 x 928M Model 4525 15000 SENSITIVE AREA Drawing 511 x 928N Model 4525 AREIAL VIEW Buffer Zone Drawing 511 x 928P Model 4525 Series PLAIN VIEW Drawing 511 x 928Q Model 4525 STAND RIGHT LEG ...

Страница 39: ...ection4 Ludlum Measurements Inc Page 4 2 December 2020 Model 4525 COMPUTER Drawing 396 x 850 Model 4525 ANCHOR BOLT TEMPLATE Drawing 511 x 997 Model 4525 ANCHOR BOLT AND CONCRETE Drawing 511 x 836 Model 4525 GEN IV LIGHTNING PROTECTION Drawing 511 x 1067 ...

Страница 40: ......

Страница 41: ......

Страница 42: ......

Страница 43: ......

Страница 44: ......

Страница 45: ......

Страница 46: ......

Страница 47: ......

Страница 48: ......

Страница 49: ......

Страница 50: ......

Страница 51: ......

Страница 52: ......

Страница 53: ......

Страница 54: ......

Страница 55: ......

Страница 56: ......

Страница 57: ......

Страница 58: ......

Страница 59: ......

Страница 60: ......

Страница 61: ......

Страница 62: ......

Страница 63: ......

Страница 64: ......

Страница 65: ......

Страница 66: ......

Страница 67: ......

Страница 68: ......

Страница 69: ......

Страница 70: ......

Страница 71: ......

Страница 72: ......

Страница 73: ......

Страница 74: ......

Страница 75: ......

Страница 76: ......

Страница 77: ......

Отзывы: