Ludlum Measurements 319110 Скачать руководство пользователя страница 7

Model 44-110-4 Windowless Gas Flow Proportional Detector 

 

 

page 2 

 

before  starting  count.    Purge  times  can  be 
adjusted  by  adjusting  the  gas  flow  rate,  but 
excessive  flow  could  loosen  contamination 
on  the  samples  and  contaminate  the  interior 
of the chamber. 
 
After count is complete, turn off toggle valve 

and open drawer. 
 
A  background  count  should  be  taken  after 
each 

source 

count 

to 

check 

for 

contamination  on  the  sample  holder  or  area 
within the O-ring. 

 
4.  CALIBRATION

  

 

CAUTION:

  Do  not  tip  sample  counter 

over,  which  could  cause  contamination 
of  the  anode  wires  and  damage  or 
destroy the instrument. 

 

For  instruments  with  separate  power 
supplies (fixed threshold), the alpha channel 
will operate at a lower voltage than the beta 
channel. 
 

4.1  Counting Instrument

  

     

Calibrated scaler instrument  

 

HV range, nominally 1700-1900 V  

 

Nominal input sensitivity: 

 

-2 

mV 

for 

single-channel 

instrument or 
 

alpha channel = -75 mV 

 

beta  channel  =  -2  mV  (with  upper 
discriminator set at 50 mV) 

 

4.2 Operating Voltage

  

   

1.

 

Set  the  scaler  count  time  for  60 
seconds. 
 

2.

 

Connect 

the 

Model 

44-110-4 

detector  and  place  the  source  in  the 
sample cavity.  Close drawer, turn on 
gas, and allow chamber to purge. 
 

3.

 

Begin  the  search  for  the  operating 
voltage  at  1600  V.  Observing  the 
count,  increase  the  HV  until  the 
desired  efficiency  is  achieved.  Note 
the HV. 
 

4.

 

Using one-minute count times, record count 
rates  for  the  desired  alpha  and/or  beta 
sources. The starting HV value should be 25 
V  less  than  the  HV  reading  obtained  in  the 
above  step.  Continue  in  25-volt  steps  until 
any  one  of  the  necessary  criteria  is  not  met 
(that  is,  alpha  or  beta  efficiency,  acceptable 
background, or “cross talk” parameters). The 
ending  HV  value  should  be  at  least  25  V 
greater  than  the  HV  reading  obtained  in  the 
above step. 
 

5.

 

Select  the  optimum  operating  voltage  that 
gives  the  greatest  alpha  and  beta  source 
efficiency and acceptable background cpm. 
 

 

4.3 Calculating Efficiency 

 
1.

 

NIST-traceable sources required. 

 
2.

 

Set HV as determined above. 

 

3.

 

Record  a  one-minute  background  and 
one-minute  source  count.  Subtract  the 
background count from the source count. 
Divide  the  net  source  count  by  the  dpm 
value  of  the  source,  times  100  for  4π 
efficiency.  
 

If  the  source  value  is  listed  in  µCi 

(activity): 

4.

 

Convert the µCi value to a dpm value by 
multiplying the microcurie value by 2.22 
x  10

6

.  Calculate  the  4π  efficiency  as  in 

the previous steps. 

 

Содержание 319110

Страница 1: ...LUDLUM MODEL 44 110 4 WINDOWLESS GAS FLOW PROPORTIONAL DETECTOR October 2020 Serial Number 319110 and Succeeding Serial Numbers ...

Страница 2: ...LUDLUM MODEL 44 110 4 WINDOWLESS GAS FLOW PROPORTIONAL DETECTOR October 2020 Serial Number 319110 and Succeeding Serial Numbers ...

Страница 3: ...anty of merchantability or fitness which extend beyond the description of the face there of If the product does not perform as warranted herein purchaser s sole remedy shall be repair or replacement at the option of Ludlum Measurements In no event will Ludlum Measurements be liable for damages lost revenue lost wages or any other incidental or consequential damages arising from the purchase use or...

Страница 4: ...Model 44 110 4 ...

Страница 5: ...Flow Proportional Detector TABLE OF CONTENTS 1 GENERAL 1 2 SPECIFICATIONS 1 3 OPERATING PROCEDURES 1 4 CALIBRATION 2 4 1 Counting Instrument 2 4 2 Operating Voltage 2 4 3 Calculating Efficiency 2 PARTS LIST 3 DRAWINGS AND DIAGRAMS 4 ...

Страница 6: ...dized aluminum tray with 2 5 cm 1 0 in diameter sample ring to allow for 2 5 or 5 1 cm 1 0 or 2 0 in Dia x L samples CONSTRUCTION aluminum housing TEMPERATURE 20 to 50 C 4 to 122 F SIZE 15 2 x 11 4 x 23 6 cm 6 x 4 5 x 9 3 in H x W x L WEIGHT 1 5 kg 3 3 lb 3 OPERATING PROCEDURES Connect the Model 44 110 4 to an appropriate counting gas by connecting tubing from a gas bottle regulator or adjustable ...

Страница 7: ...source in the sample cavity Close drawer turn on gas and allow chamber to purge 3 Begin the search for the operating voltage at 1600 V Observing the count increase the HV until the desired efficiency is achieved Note the HV 4 Using one minute count times record count rates for the desired alpha and or beta sources The starting HV value should be 25 V less than the HV reading obtained in the above ...

Страница 8: ...RECPT UG706 U C 4478 011 1 EA KNOB POINTER MS91258 2P2B 8 6608 1 EA O RING 2 229 16 8286 1 EA O RING 2 012 16 8302 4 EA FEEDTHRU 1115 07 0519 18 9353 1 EA SPRING PLUNGER 10 32 18 9365 4 EA MOUNT ISOLATION 8 32 X 1 2 BLK 21 8922 2 EA FTG 1 8 HOSE MH1332 1 4 HEX 21 9305 1 EA VALVE MTV 2P 21 9754 1 EA M 44 10 43 17 SMPL DRWR 7142 001 06 1 EA M 43 10 PLANCHET HOLDER 7142 001 07 1 EA M 43 10 43 17 SHAF...

Страница 9: ...ional Detector page 4 DRAWINGS AND DIAGRAMS Model 44 110 4 Assembly Overview Drawing 142 x 274 Model 44 110 4 Assembly Step 1 Drawing 142 x 274A Model 44 110 4 Assembly Step 2 Drawing 142 x 274B Model 44 110 4 Assembly Step 3 Drawing 142 x 274C ...

Страница 10: ......

Страница 11: ......

Страница 12: ......

Страница 13: ......

Отзывы: