Ludlum Measurements 26 Скачать руководство пользователя страница 2

7 - Setup Parameter Order 

Setup Parameter 

Default 

Value 

Notes 

Editing 

Unit of Measurement 

cpm 

cpm 
cps 

MODE to select  value 

Response Time 

0 = Auto Response Rate Mode 
1-60 = Fixed Response Time (seconds) 

OK to select digit 
MODE to select  value 

Auto-Response Rate 
Mode 

 

Count Rate 

F = Fast 

(seconds)

S = Slow 

(seconds) 

Less than 3 kcpm 

10 

Between 3 - 4 kcpm  

Between 4 - 6 kcpm  

Between 6 – 12 kcpm 

More than 12 kcpm 

     

MODE to select  value 

Count Rate Alarm 

0-1999 selectable range 

OK to select digit 
MODE to select  value 

Scaler Time 

1 minute 

0 = continuous count 
1 sec to 20 minute range (max value = 19 min 60 sec) 

OK to select digit 
MODE to select  value 

Operational Modes 

0

 : RATE, MAX, SCALER  

1:  

RATE, MAX  

2: 

RATE,SCALER 

3: 

RATE  

MODE to select  value 

 

8 – Protective Mode Switch 

An internal DIP (dual inline position) switch can be set to either permit editing parameters (see section 6 – Initiating 
Setup) or to Protect against any editing by users.  To gain access to the DIP switch, remove the batteries, then unscrew the 
4 screws holding the instrument together so the instrument halves become separated.   

Once separated, gently move the GM pancake 
detector from its holder over to one side while 
leaving the wire intact.  With the detector out of 
the way, you will see the dual switches as 
indicated in the picture.   

When both switches are slid to the LEFT 
position it is in the Edit Mode (default). 

When the bottom switch is in the Right Position, 
it is in the Protect Mode. 

 