How to Use SDHT Pass Through Testing
When using Pass Through Testing, a target value (either in Force or
Length), must be preset. If using a Two Point mode then two target values
must be preset.
There are two limitations when using Pass Through Testing:
•
First, Pass Through Testing data storage is finite. When performing
a test, the amount of storage used will be displayed. When storage
reaches 100%, the test will finish even if the target has not been
passed.
•
Second, to get the best readings the rate of force change must be
below a certain level. This changes with the rate of the part being
tested. If the tester detects the rate of change of the force is too high
a message will be displayed suggesting to slow down.
Setup
Each testing mode,
Manual
or any of the
Two Point with Rate
modes, has its own target values. The mode to be used must
be selected using the
MODE
button
before
accessing the
Pass
Through Testing
setup menu.
1) Press the
OPTION
button until the screen shows “
Pass
Through
”.
2) Press
F1
to enter the “Setup” menu.
3) Using the
F1
button select the target measurement:
Force
or
Length
.
4) Press
OPTION
to go to the next menu.
5) Using the
F1
and
F2
buttons, enter the target value.
6) Press
OPTION
to go to the next menu.
7) If the current mode is a
Two Point
mode then a second target
can be entered.
8) When all target points are set press the
OPTION
button to
scroll back to the 1
st
target point.
9) Press the
ON/CLEAR
button to exit back to the primary
display. If you want to immediately start the
Pass Through
Testing
mode, press the
SEND
button to move up a menu
level and press
F2
to
Start
using
Pass Through Testing
.
Using Pass Through Testing – Continued on Page 2