MANUAL

LCC 100

Operang Modes and Operaon

1) CTRL (control)

The CTRL mode is used for calibraon of pressure sensors, switches and other pressure equipment.
The pressure pump is acve and regulate the corresponding target pressure. The internal sensor
displays the actual value on the LCD.

• Connecon

Pressure devices have to be connected to the + and the - port of the calibrator. Absolute pressure
devices have to be connected to the + port.

• Se ng

- Press the MENU bu on. Confirm MODE with OK and press UP/DOWN unl CTRL is shown on the

display. Confirm with OK bu on.

- Set the upper limit of the range in the submenu RANGE (UP/DOWN and 0%/100%), confirm with

OK.

- Set of the unit in the submenu UNIT (UP/DOWN), confirm with OK.
- Set the step sequence (in %) in the submenu STEPS (UP/DOWN and 0%/100%) and confirm with

OK.

- MENU bu on to exit the submenu.

• Use

The unit is located at exit from the menu in VENT mode, ie the sensor of the calibrator is vented. On
the middle of the top, the target pressure is displayed. On the bo on of the display, the % value is
shown. When leaving the menu, this value is 0%. Switch between target pressure and % value with
the OK bu on.
- The % value can be changed with the UP/DOWN (in the fixed step sizes) and using the 0% / 100%

bu ons.

- The target pressure can only be changed with UP/DOWN. With 0%/100% to the appropriate place,

the target pressure be jumped.

• Example

Target pressure: By pressing the OK bu on, you can switch between the % and target pressure value.
Press OK buton unl the value (top center display) outlined in bold. Press the 0% or 100% bu on to
jump to the selected locaon. Choose the required value with the UP and DOWN bu ons. Push the
OK bu on to exit target pressure. The marker moves to the % value (middle of the display).

When the calibrator displays a stable value, the data can be recorded:

- Nominal value: right below
- Unit: right above
- Actual value: on the display of the connected device
- Reference value: middle of the display

You can make a manual zero of the internal sensor in the CTRL mode. Connected devices can be leak
testet:

- Manual auto zero: Push the ZERO bu on, the internal sensor is get to zero.

- Leak-Test: Push the LEAK bu on to switch off the pressure cell. The pressure is held. Control the

value via the display of the calibrator. Push the LEAK bu on again to stop the mode.

4

Содержание LCC100

Страница 1: ...e Controller LCC 100 LR Cal DRUCK TEMPERATUR Leitenberger GmbH Bahnhofstr 33 D 72138 Kirchentellinsfurt GERMANY Tel 49 0 7121 90920 0 Fax 49 0 7121 90920 99 DT Info Leitenberger de www druck temperatu...

Страница 2: ...g mode Applica on Func on CTRL control Calibra on of sensors Pressure is built up and maintained and other devices Target pressure may be percentage or absolute MESS measure Measurement of The calibra...

Страница 3: ...Start the AUTO mode with the OK bu on In CTRL mode you can use the bu on to change the value from 0 to 100 In submenu for se ng the decimal The display you can always check the ba ery status and the...

Страница 4: ...e top the target pressure is displayed On the bo on of the display the value is shown When leaving the menu this value is 0 Switch between target pressure and value with the OK bu on The value can be...

Страница 5: ...ubmenu SETTINGS Push the UP DOWN bu on to get the item SETTINGS and con rm with OK Choose the item ZERO in the submenu and con rm with OK Via the item AUTO ZERO you can de ac vate or ac vate the auto...

Страница 6: ...to sec ons Further se ngs of AUTO mode can be set in the submenu CONFIGURATION Cycles Set the number of cycles that have to be realized in the AUTO mode T Start Wait for vent of the internal sensor St...

Страница 7: ...alibrator can be evaluated and recorded Exit of AUTO mode with the MENU bu on VENT is shown le above in the display 4 LEAK leakage In the CTRL mode it is possible to check connected pressure device fo...

Страница 8: ...ayed The following se ngs can be made AKTIV if a interface is required choose of the individual interface USB RS232 BAUD choose of the correct baud rate 1200 2400 4800 9600 14400 19200 28800 56000 576...

Страница 9: ...the dwell me starts running ats 1 10000 CR Start delay phase 2 1 10000 Time in seconds O On Interface output o 0 1 CR Output status informa on via interface 0 1 Output format of con nuous status info...

Страница 10: ...1 ps 10 110 CR Enter set point targe n percent 10 110 Set point target in pu CR Increase set value by increment in Step UP Menu se ngs saz 0 1 CR Set Auto Zero in modes Measure MESS and CTRL 0 1 sbr 0...

Страница 11: ...fp 1 99999 CR Pitot tube factor 0 99999 Pitot tube factor nondimensional 10000 sfu 0 3 CR Flow rate unit 0 meter per second m s 1 kilometer per hour km h 2 f per minute fpm eet 3 miles per hour mph sm...

Страница 12: ...MANUAL LCC 100 12 szi 1 60 CR Zero se ng interval 1 60 Time in minutes szm 0 1 CR Zeroing during quit of the menu 0 1 O On DRUCK TEMPERATUR Leitenberger GmbH Bahnhofstr 33 D 72138 Kirchentellinsfurt...

Отзывы: