20
21
diagnostics
3 diagnostics
>
3.2 equiPMent PerForMance (cont.)
Problem
Probable causes
Solutions
Poor suction
from the main
massage head.
Wear on the head on the outside bevel
of the treatment chamber
Change the head.
The head has exceeded 1000 hours
of use.
Change the head or keymodule.
The flaps may be stuck due to tissue
desquamation.
Detach and clean the flaps.
The hose has a hole.
Change the hose.
Test the seal from the onboard test menu.
Suction loss on one
of the heads
Tissue desquamation has formed
plugged up a “plug and care” hose
connection to the head or machine.
Clean the hose connections.
Test the seal from the onboard test menu.
Painful treatment for
the patient. Ineffective
treatment.
Jerking movement.
Excessive suction in the heads. Loss of
rhythmicity in the suction sequence.
Change the pneumatic control assembly.
Change the control board.
Test the control from the onboard
test menu.
cont .
>
3.3 general critical ProBleMS
Problem
Probable causes
Solutions
Machine not operating.
Wrongly positioned power cord
Check the connection to the machine
and to the power outlet.
Press the power switch.
No light on the switch.
The machine is not plugged in
properly. If the cord is too long or to
short, it creates too much resistance.
Check that the socket has power.
Plug it in and switch on the power.
Black screen on the
panel, voltage indicator
and rear green
indicator lit
Panel fails to start up
Replace the panel.
No back-lighting
diagnostics
3 diagnostics
3.1 noiSeS
3.2 equiPMent PerForMance
Problem
Probable causes
Solutions
Pump is too loud:
clicking.
Absorption of a foreign body
Check for a foreign body.
Loud pump: (foghorn)
Defective silencer mechanism
Replace the silencer.
Pump vibration noise
Loose attachment
Tighten the pump attachment.
Whistling sound in the
filter compartment
Loose filter
Tighten the filter.
Test the pneumatic seal from
the onboard test menu.
Whistling sound
in the hose
Cut hose
Replace the hose.
Test the pneumatic seal from
the onboard test menu.
Problem
Probable causes
Solutions
Poor performance
by a pump
Blocked flap, presence of
a foreign body
Replace the pump.
Test the pump from the
onboard test menu.
Pump stops and
does not start back up,
weak suction
on the head
Thermal overload
Check the pump’s coolant.
Pump connection problem
Check the pump power connection,
and check the cable in the
pump compartment.
Identify the defective pump using
the onboard test menu.
Weak or no suction
in the heads
The filter is not completely screwed in.
There is a leak in the threading.
Check that the filter is positioned correctly.
No filter
Put in a new filter.
Missing filter compartment seal
Put in a new seal.
Test the seal from the onboard test menu.
Stopped pump
See above.
Weak suction from the
auxiliary heads.
The hose has a hole.
Change the hose.
Test the seal from the onboard test menu.
cont .
>