3 Product description
LM
T 6
86
51
c 0
2/2
02
2
EN
|
11
3.8 Data management/
compatibility
3.8.1 Saving and transmitting therapy
data
Therapy data for the previous 30 therapy days (24 hours/day)
are saved in the device. Pressure, flow and volume are saved
at 20 Hz, all other recorded values at 1 Hz.
Statistical data for the previous 12 months are saved in the
device.
A file in edf format is created for every day saved.
If you plug USB flash drive LMT 31414 into the device, the
therapy data saved in the device will be transmitted to the
flash drive in the form of edf files.
The therapy data saved on the USB flash drive can be read
into and displayed in the prismaTS software.
3.8.2 Updating the firmware
In order to perform a firmware update, plug a USB flash drive
with an update file (one version higher than the current
version) into the device and confirm that the update should
go ahead.
The device configuration is retained following the update.
3.8.3 Setting up a connection to the
LUISA app
The LUISA app is an app on a mobile terminal. The device can
be connected to the LUISA app
Material damage if incorrectly configured!
If trolley 2.0 is not used properly, it may tip over or be
damaged.
Use the circuit holder only for the circuit.
Use the water bag holder only for the refill unit of the
active humidification system.
Use trolley 2.0 only on a maximum incline of 10°.
Ensure that the total weight of trolley 2.0 when fully
equipped is < 25 kg.
Before moving the trolley: Put the circuit holder in the
folded-away position.
Anyone who integrates medical devices or medical
software products in an IT network or installs them
on a PC or integrates devices or software products in
a medical IT network or installs them on a PC is
responsible for complying with IEC 80001-1.
According to IEC 80001-1, the owner/operator is
responsible for the risk management of any
interactions in medical IT networks. Please note that
the manufacturer does not accept any warranty or
liability for interactions between system components
in an IT network.