4 Preparation and operation
LM
T 6
86
51
c 0
2/2
02
2
EN
|
15
3. Press the
Ok
key to start calibration.
4. If calibration is successful, press the
Finish
key.
If calibration is not successful, follow the instructions in
the display and eliminate the faults.
5. Continue O
2
supply.
The FiO
2
cell is continuously emptied as a result of contact
with oxygen. If the FiO
2
cell is almost or completely empty,
an alarm message will appear. The FiO
2
cell is fitted and
replaced by the specialist dealer.
4.7 Pairing device with LUISA
app
The LUISA app is an app on a mobile terminal which you can
use to read off the patient’s therapy data.
1. Activate the
Bluetooth
function in the
System
>
Device settings
>
Connectivity
menu.
2. Select the entry
Add new device
in the
Device list
menu.
3. Download the app onto a mobile terminal and follow
the instructions in the app.
After pairing, the app will recognize the Bluetooth
connection of the device. The pairing does not then need to
be performed again. The saved pairing can be deleted in the
LUISA app.
To stop the alarm occurring, you can deactivate the
cell in the
System
>
FiO
2
cell
menu.