7 Alarms
LM
T 6
86
51
c 0
2/2
02
2
EN
|
27
Error FiO
2
cell
770
FiO
2
cell defective.
Contact your specialist dealer. Have FiO
2
cell
replaced.
No FiO
2
cell
771
No FiO
2
cell fitted.
Contact your specialist dealer. Have FiO
2
cell
fitted.
FiO
2
cell empty
773
FiO
2
cell empty.
Contact your specialist dealer. Have FiO
2
cell
replaced.
Blower temperature
high
789
Blower temperature too high. Cooling
air filter blocked.
Cool device immediately or therapy will end.
Check cooling air filter. If necessary: Have
cooling air filter replaced by specialist dealer.
SpO
2
signal weak
SpO
2
signal weak
792
790
SpO
2
sensor not connected to the
finger correctly.
Check connection with the finger. If alarm
persists: Contact your specialist dealer.
Signal interfered with by nail varnish or
contaminants.
Remove nail varnish. Clean finger.
SpO
2
sensor removed
791
SpO
2
sensor removed.
Reconnect SpO
2
sensor.
If the alarm persists: Replace SpO
2
sensor.
SpO
2
cable removed
793
SpO
2
cable removed.
Reconnect SpO
2
cable.
Therapy ended
794
Device is switched off.
Switch device back on.
Faulty circuit
795
Single circuit with valve selected in
menu but double circuit connected.
Change circuit or have the medical professional
or the specialist dealer set the connected circuit
in the menu.
Have the medical professional check the
settings.
Leakage circuit selected in menu but
single circuit with valve connected.
Change circuit or have the medical professional
or the specialist dealer set the connected circuit
in the menu.
Have the medical professional check the
settings.
Change circuit or have the medical professional
or the specialist dealer set the connected circuit
in the menu.
Circuit defective.
Check circuit and the correct fit of circuit. If
necessary, replace circuit.
Re-inhalation
796
Valve does not open in exhalation
(medication has caused it to stick, for
example).
Check circuit and the correct fit of circuit. If
necessary, replace circuit.
Patient's re-inhalation volume
excessive at high frequency.
Disconnection valve
control pressure
798
Single circuit with valve selected in
menu: Valve control tube not
connected correctly or not connected
at all.
Check valve control tube and connect correctly.
Single circuit with valve selected in
menu but leakage circuit connected.
Change circuit or have the medical professional
or the specialist dealer set the connected circuit
in the menu.
Blower overheated
799
Blower has overheated.
Therapy will end.
Allow device to cool down.
DISPLAY
CODE
CAUSE
ACTION