Ventilation modes and parameters | A 5
Instructions for Use elisa 300-500 | Supplement A |
SW 2.09.x
| Rev. 11
95
A 5.4
Ventilation parameters
Different parameters are available to set the ventilation modes:
O
2
Inspiratory O
2
concentration
PEEP
Positive end-expiratory pressure
Insp. Ramp
Inspiratory time ramp between PEEP and Pinsp
PS Ramp
Time ramp between PEEP and PS
Pinsp
Inspiratory pressure
PS
Pressure support
Tinsp
Inspiratory time,
inspiratory/expiratory ratio I:E is also shown
PS TImax
Maximum duration of inspiratory phase with pressure support
Exp. trigger
Expiratory trigger with pressure-controlled ventilation
Percentage of the trigger expectation window
for the synchronised switch to expiration
PS Endflow
Stop criterion of inspiratory phase with pressure support
Rate
Respiratory rate
Trigger
The ventilator detects the patient's attempts to breathe in via the trigger.
Trigger sensitivity can be individually adapted to the patient. The lower the
set value the more sensitive the ventilator is to patient's attempts at inhaling.
At very low settings, even the elasticity of the ventilation tubes might cause a
trigger signal (auto-triggering).
Vt
Tidal volume
Byflow
As bias flow, the Byflow provides the set trigger flow and simultaneously pre
-
sents a kind of “flow reservoir” for the spontaneously breathing patient. The
Byflow function is available in all ventilation modes. The setting can be in
-
creased if the dynamic behaviour of the device is unsatisfactory during or after
triggering when using endotracheal tubes with a deflated cuff, or if rebreathing
of CO
2
is to be prevented when a ventilation helmet is used.
Pmax
A pressure limit can be set in order to rule out excessively high pressure
values in case of an acute worsening of the mechanics of breathing. No further
pressure adjustments are carried out if the pressure limit has been reached.
The “Pressure limitation” alarm is triggered. The Vtmin alarm limit must be
set in order to be able to recognise that the set tidal volume has not been
reached.
For volume-controlled modes, the pressure limitation Pmax can be deactivat
-
ed on the configuration level.
Pmin
In pressure-controlled or pressure-supported ventilation modes with speci-
fied volumes, Pmin represents the lower pressure adaptation window. If the
required pressure drops below this setting, the automatic pressure adaptation
is suspended. To prevent unintended high tidal volumes, it is necessary to set
the Vtmax alarm limit.
I-Flow
Inspiratory gas flow rate
Tplateau
Duration of the plateau during the inspiratory phase (No-flow phase)
Flow support
Extent of the resistance compensation
Содержание elisa 300
Страница 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Страница 10: ...8 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 0 Contents This page is intentionally left blank ...
Страница 18: ...2 Use 16 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Страница 102: ...7 Operation 100 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Страница 246: ...A 3 Functions 30 Instructions for Use elisa 300 500 Supplement A SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Страница 362: ...12 Instructions for Use Supplement B CO2 Sensor LeoCap Rev 04 elisa 300 elisa 500 elisa 600 elisa 800 ...
Страница 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Страница 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Страница 423: ...Instructions for Use Supplement D Pulse Oximetry and LeoClac Rev 01 17 This page is intentionally left blank ...
Страница 425: ...elisa 300 elisa 500 elisa 600 elisa 800 elisa 800 VIT External Communication Instructions for Use Supplement E ...
Страница 432: ...8 Instructions for Use Supplement E External Communication Rev 02 Rear view of elisa 600 Rear view of elisa 800 ...
Страница 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Страница 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...