A 4 | Weaninganalyzer
®
86
Instructions for Use elisa 300-500 | Supplement A |
SW 2.09.x
| Rev. 11
On the monitoring screen, adapt the target values to the patient’s clinical situation:
An adaptive security ventilation mode is active during the SBT manoeuvre. This
feature supplements the ventilation within the cycle time in the event of missed
spontaneous breaths; audible alarms are not released in these situations. Before
activating the SBT, adjust the cycle time and the required pressure difference (Pin
-
sp SBT) to the clinical situation. When necessary, pressure support (PS SBT) can
be added during an SBT manoeuvre.
Beginning and end of the most recent SBT are displayed in the upper left hand
corner of the SBT Workbench.
A 4.4.1
SBT procedure
*
The rate for adaptive
security ventilation is
calculated by dividing
60
by the
cycle time
Perform SBT
Check tube compensation setting
Activate tube compensation
On monitoring screen, adapt target values to clini
-
cal situation
Adjust SBT duration (=virtual extubation)
Activate P0.1 measurement
(recommended: automatic start of P0.1, interval of
5 minutes)
Adjust adaptive security ventilation
*
:
- Cycle time
- Pinsp SBT
- PS SBT if applicable
Activate SBT (= ON)
SBT starts automatically when all target values
remain in the target range for 5 minutes.
Содержание elisa 300
Страница 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Страница 10: ...8 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 0 Contents This page is intentionally left blank ...
Страница 18: ...2 Use 16 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Страница 102: ...7 Operation 100 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Страница 246: ...A 3 Functions 30 Instructions for Use elisa 300 500 Supplement A SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Страница 362: ...12 Instructions for Use Supplement B CO2 Sensor LeoCap Rev 04 elisa 300 elisa 500 elisa 600 elisa 800 ...
Страница 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Страница 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Страница 423: ...Instructions for Use Supplement D Pulse Oximetry and LeoClac Rev 01 17 This page is intentionally left blank ...
Страница 425: ...elisa 300 elisa 500 elisa 600 elisa 800 elisa 800 VIT External Communication Instructions for Use Supplement E ...
Страница 432: ...8 Instructions for Use Supplement E External Communication Rev 02 Rear view of elisa 600 Rear view of elisa 800 ...
Страница 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Страница 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...