A 3 | Functions
34
Instructions for Use elisa 300-500 | Supplement A |
SW 2.09.x
| Rev. 11
A 3.3.1.3
Postoxygenation phase
If a closed breathing circuit is detected, the suction phase is terminated and venti-
lation in the post-oxygenation phase starts. Ventilation will also start if the patient
is still not disconnected after 10 seconds.
Alternatively, the ventilator switches to the post-oxygenation phase once the
suction phase has been completed and starts ventilation. If this is not possible
because the breathing circuit is still open, a corresponding alarm is triggered.
Once ventilation is resumed, the ventilation parameters and alarm limits of the
pre-oxygenation phase are applied. Even a modified O
2
value is maintained but it
can be changed at any time.
The post-oxygenation phase is terminated after 2 minutes. The O
2
value is auto-
matically reset to the value which was set prior to starting the automatic suction
routine. The automatic suction routine switches off automatically.
A 3.3.2
Suction with closed suction systems
When performing the suction routine with a closed suction system, the set oxy
-
gen concentration is applied as complementary oxygenation for 120 seconds, the
byflow is adapted, and at the same time some of the audible alarms are disabled.
Remark
The BiLevel ventilation mode should be selected for suction with
closed suction systems because negative pressures may occur or
control algorithms may be altered in other modes.
Remark
The previously selected settings are retained during closed suc-
tion. The O
2
supply is increased during the manoeuvre.
A 3.4
Cuffscout
Remark
The Cuffscout function is not available for elisa 300.
A 3.4.1
Device setup
Two different Cuffscout lines are available and either type can be used.
Warning
Only the Cuffscout connecting lines included in the accessory list
are approved for use with the Cuffscout function (see illustration).
Содержание elisa 300
Страница 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Страница 10: ...8 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 0 Contents This page is intentionally left blank ...
Страница 18: ...2 Use 16 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Страница 102: ...7 Operation 100 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Страница 246: ...A 3 Functions 30 Instructions for Use elisa 300 500 Supplement A SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Страница 362: ...12 Instructions for Use Supplement B CO2 Sensor LeoCap Rev 04 elisa 300 elisa 500 elisa 600 elisa 800 ...
Страница 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Страница 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Страница 423: ...Instructions for Use Supplement D Pulse Oximetry and LeoClac Rev 01 17 This page is intentionally left blank ...
Страница 425: ...elisa 300 elisa 500 elisa 600 elisa 800 elisa 800 VIT External Communication Instructions for Use Supplement E ...
Страница 432: ...8 Instructions for Use Supplement E External Communication Rev 02 Rear view of elisa 600 Rear view of elisa 800 ...
Страница 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Страница 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...