8 | Measurements and graphics
106
Basic Instructions for Use elisa 300-500
|
SW 2.09.x
|
Rev. 11
Depending on the settings on the configuration level, further views with the pre
-
configured measurements will be available.
Remark
The expiratory time constant is a measure of the complete ex
-
piration while avoiding an auto-PEEP situation. For this measure-
ment, the expiratory flow is determined at 75% of the expiratory
vital volume and the necessary expiration time is calculated.
The following standard values are applicable for the necessary
expiration time:
1 x RCexp 63%
2 x RCexp 86.5%
3 x RCexp 95%
4 x RCexp 98%
8.5
Fastwean
Measurement values and indexes relevant for the weaning procedure can be
monitored in the “Fastwean” weaning window. Values in the “safe window” are
displayed in green, values outside the limits are displayed in yellow (the displayed
values as well as the limits may be different, depending on the configuration):
Yellow
Green
Yellow
P0.1 (mbar)
< –5
-5–0
-
RR spont. (/min)
< 9
9–16
> 16
VTe spont.
(ml)
< 200
200–450
> 450
RSBI
-
0–105
> 105
Touch the upper limit or the lower limit and adjust the corresponding value with
the selector.
These limits only apply to Fastwean, they do not affect other display elements,
such as Fastprotect.
The general alarms are also not affected.
Содержание elisa 300
Страница 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Страница 10: ...8 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 0 Contents This page is intentionally left blank ...
Страница 18: ...2 Use 16 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Страница 102: ...7 Operation 100 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Страница 246: ...A 3 Functions 30 Instructions for Use elisa 300 500 Supplement A SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Страница 362: ...12 Instructions for Use Supplement B CO2 Sensor LeoCap Rev 04 elisa 300 elisa 500 elisa 600 elisa 800 ...
Страница 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Страница 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Страница 423: ...Instructions for Use Supplement D Pulse Oximetry and LeoClac Rev 01 17 This page is intentionally left blank ...
Страница 425: ...elisa 300 elisa 500 elisa 600 elisa 800 elisa 800 VIT External Communication Instructions for Use Supplement E ...
Страница 432: ...8 Instructions for Use Supplement E External Communication Rev 02 Rear view of elisa 600 Rear view of elisa 800 ...
Страница 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Страница 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...