56
PM620_4c 07/2015
1.1 Methods
21. Close the vial tightly with the cap and swirl several times
until the tablet is dissolved.
22. Place the vial in the sample chamber making sure that
the marks are aligned.
23. Wait for a
reaction period of 2 minutes
.
24. Press
TEST
key.
Record the displayed test result (C).
25. Remove the vial from the sample chamber.
26. Add
one DPD ACIDIFYING tablet
straight from the foil
to the same water sample and crush the tablet using a
clean stirring rod.
27. Wait for a
reaction period of 2 minutes
.
28. Add
one DPD NEUTRALISING tablet
straight from
the foil to the same water sample and crush the tablet
using a clean stirring rod.
29. Close the vial tightly with the cap and swirl several times
until the tablets are dissolved.
30. Place the vial in the sample chamber making sure that
the marks are aligned.
31. Press
TEST
key.
Record the displayed test result (D).
Zero accepted
prepare Test
press TEST
Zero accepted
prepare Test
press TEST
Содержание PM 620
Страница 1: ...Instruction manual Photometer System P 620 GB...
Страница 2: ......
Страница 4: ......
Страница 14: ...6 PM620_4c 07 2015...
Страница 15: ...7 PM620_4c 07 2015 Part 1 Methods...
Страница 19: ...11 PM620_4c 07 2015...
Страница 93: ...85 PM620_4c 07 2015...
Страница 118: ...110 PM620_4c 07 2015...
Страница 119: ...111 PM620_4c 07 2015 Part 2 Instrument Manual...
Страница 122: ...114 PM620_4c 07 2015...
Страница 169: ...161 PM620_4c 07 2015 Part 3 Enclosure...
Страница 177: ...169 PM620_4c 07 2015...
Страница 178: ...170 PM620_4c 07 2015...
Страница 179: ...171 PM620_4c 07 2015...