29
CheckitDirect_Cl-Br-pH-CyA-TA-CaH_2 09/2010
GB
Methods
RESULT
RESULT
Chlorine
with Tablet
0.01 – 6.0 mg/l
a) free Chlorine
Fill a clean vial (24 mm Ø) with
10 ml of the water sample
and
perform zero calibration (see “Operation”).
Remove the vial from the sample chamber and empty it, leaving a few
drops remaining in the vial.
Add
one DPD No. 1 tablet
straight from the foil and crush the tablet
using a clean stirring rod.
Add the water sample to the 10 ml mark. Close the vial tightly with the
cap and swirl gently several times until the tablet is dissolved.
Place the vial in the sample chamber making sure that the marks
are aligned.
Press the [ZERO/TEST] key.
The method symbol flashes for approx. 3 seconds.
The result is shown in the display in mg/l free Chlorine.
b) total Chlorine
Add
one DPD No. 3 tablet
straight from the foil to the same water
sample and crush the tablet using a clean stirring rod. Close the vial
tightly with the cap and swirl gently several times until the tablet is
dissolved.
Place the vial in the sample chamber making sure that the marks
are aligned.
Wait for a reaction period of 2 minutes.
(Countdown can be activated, see page 27)
The method symbol flashes for approx. 3 seconds.
The result is shown in the display in mg/l total Chlorine.
c) combined Chlorine
combined Chlorine = total Chlorine – free Chlorine
Tolerances:
0 – 1 mg/l: ± 0.05 mg/l
> 1 – 2 mg/l: ± 0.10 mg/l
> 2 – 3 mg/l: ± 0.20 mg/l
> 3 – 4 mg/l: ± 0.30 mg/l
> 4 – 6 mg/l: ± 0.40 mg/l
Notes:
see page 28
0.0.0
CL
CL
!
Zero
Test
Zero
Test
Содержание CheckitDirect
Страница 17: ...17 CheckitDirect_Cl Br pH CyA TA CaH_2 09 2010 DE Methoden ...
Страница 37: ...37 CheckitDirect_Cl Br pH CyA TA CaH_2 09 2010 GB Methods ...
Страница 57: ...57 CheckitDirect_Cl Br pH CyA TA CaH_2 09 2010 FR Méthodes ...
Страница 77: ...77 CheckitDirect_Cl Br pH CyA TA CaH_2 09 2010 IT Mètodi ...
Страница 97: ...97 CheckitDirect_Cl Br pH CyA TA CaH_2 09 2010 ES Métodos ...
Страница 103: ...103 CheckitDirect_Cl Br pH CyA TA CaH_2 09 2010 ...