Rev B
32
synthesizer count value recorded on the label. If the synthesizer count values do not match
then perform the remainder of this procedure up to step 14 then repeat this procedure
beginning at step 6. In the event that the values still do not match contact LOTEK Engineering
Inc. for assistance.
12.
The message ‘Set decode threshold?’ will be displayed. If the label containing the LO and IF
information also contains a 2 digit threshold variable, answer yes, then enter the 2 digit
number. Otherwise answer no. (This entry will not affect the sensitivity of the receiver, but
allows the optimum gain value for an automatic station, established by the factory, to apply
universally to any receiver).
13.
The message ‘Test extended memory?’ does not apply to all versions. If the message appears,
there is no need to perform a memory test unless a problem is suspected. Answer yes or no
accordingly.
14.
When all the above information has been entered the receiver will ‘boot’ normally and the
following information will be displayed: version information, date and time followed by the
‘command environment’ display of frequency and gain.
15.
For receivers equipped with dual bank firmware versions
W30SAT, W31, W32
, it is necessary
to do the following:
-Press <SHIFT F0> which is normally used for entering the Code_Log menu.
-The message "NewProm.." will be displayed (the same message, which would show up
after hitting <SHIFT> F1, which initiates the New Prom procedure)
-Confirm the first menu dialogue with Yes. It is not necessary to further proceed with the
initialization.
For a receiver equipped with dual bank version firmware
W40
, it is necessary to do the
following:
-Press <SHIFT> F3 (normally used to toggle between the environments)
-The message NewProm will be displayed (the same message, which would show up after
hitting <SHIFT> F1, which initiates the New Prom procedure)
-Confirm the first menu dialogue with Yes. It is not necessary to further proceed with the
initialization.
16.
If the firmware belongs to the
W16
family, simply start Code_Log with the option "Initialize"
and escape from it afterwards. It is not necessary to enter a frequency in the frequency table.
17.
Examine the receiver to make sure that no wires and cables are pinched or stressed. Secure
the cover of the shield box then secure the bottom cover of the receiver.
18.
Verify receiver operation using a new tag. The tag frequency should produce the highest
power. This can be verified by shifting the frequency up and down from the tag frequency by
1 KHz using the ‘UP-ARROW’ and ‘DOWN-ARROW’ keys. Note that the power decreases
and the pitch of the tag signal changes. If there is a large discrepancy repeat this step with a
different tag. If the results are the same then contact LOTEK Engineering Inc. for assistance.
*
The synthesizer count value is a number generated by the processor which represents the
timebase the receiver is operating on. If the number displayed does not match the number