V-Link
®
-200
User Manual
7.1.3
Manufacturer Calibration
NODE: V-Link-200 ,18 bit (262144 A/D values)
CHANNEL TYPE: differential analog input, 0 to 5 V dc input range
SENSOR TYPE: pressure transducer, voltage output, positive going
SENSOR PARAMETERS:
From the manufacturer calibration sheet included with the sensor;
sensor range: 0-250 psi
sensor zero load output: 0.0032 V dc
sensor full scale output (FSO) with 10V excitation: 86.07 mV
From the application parameters;
sensor excitation in application: scaled to 4.096 V
DESIRED OUTPUT: engineering units (EU), psi
CALCULATIONS:
Because the sensor will be powered from the node with 4.096 V, and the sensor manufacturer
calibrated it a 10 V, the manufacturer full scale output (FSO) value needs to be scaled to 3V.
(4.096 V/10 V) * 86.07 mV = 35.25 mV
Select a gain and offset scale value appropriate for the sensor. (
See Differential Input Gain and
).
The closest gain setting that accommodates 35.25 mV is +/- 39.0 mV. Using a lesser value would
exceed the input voltage capacity of the node when the sensor is at higher pressures. For more
information about gain values and associated input ranges,
See This table lists the gain settings
Multiply the sensor FSO by the gain setting to get the sensor voltage after amplification. For this
example, the range is 39.0 mV for a gain of 64.
64 * 35.25 mV = 2.256 V
64
Содержание V-Link-200
Страница 1: ...LORD USER MANUAL V Link 200 Wireless 8 Channel Analog Input Sensor Node ...
Страница 12: ...V Link 200 User Manual Figure 4 Viewing Node Diagnostic Data 12 ...
Страница 15: ...V Link 200 User Manual 15 ...
Страница 22: ...V Link 200 User Manual 22 ...
Страница 35: ...V Link 200 User Manual Figure 31 DIN Rail Mount Option 35 ...
Страница 45: ...V Link 200 User Manual Figure 36 Half and Quarter Bridge Wiring 45 ...
Страница 76: ...V Link 200 User Manual 9 Troubleshooting 9 1 Troubleshooting Guide 76 ...