6.7.2
Differential Channel Raw Voltage
To set the V- Link- 200 to show the raw voltage that is being applied (in mV) to channels 1 - 4
(differential channels), enter the slope and offset for the input range being used. (
)
Input Range
[Gain]
Slope
Offset
±156 mV
[16]
0.0012207031
-156.25
±78.1 mV
[32]
0.0006103516
-78.13
±39.0 mV
[64]
0.0003051758
-39.06
±19.5 mV
[128]
0.0001525879
-19.53
±9.76 mV
[256]
0.0000762939
-9.77
±4.88 mV
[512]
0.0000381470
-4.88
±2.44 mV
[1024]
0.0000190735
-2.44
±1.22 mV
[2048]
0.0000095367
-1.22
Table 5 -
Raw Voltage Output
The formula to derive the above raw voltage output:
Slope = 5120mv ÷ (262144 * gain)
Offset = -(5000mV ÷ (2 * gain)) (Assuming balance to mid-scale is perfect)