6.10
Thermocouples
Thermocouples can be used on the differential input channels by simply adding a high-impedance
resistor to the input. An example circuit is shown in
Figure 40 - Connecting a Thermocouple
.
Thermocouples should be calibrated by first selecting the appropriate baseline offset range, output
range, or gain and then applying know loads and calculating the slope and offset values. Using water
as the known load medium (submerging the thermocouple in ice and hot water baths) is a simple
method that can be used for calibration.
Figure 40 - Connecting a Thermocouple
52