LORD AP5103 Series Скачать руководство пользователя страница 5

CALIBRATION

Calibration of the AP5103 is accomplished by the conventional shunt technique, using an internally 

installed calibration resistor, or via the "dead-weight" method.

Calibration Resistor.   If a fixed resistor is shunted across one arm of a strain gage bridge, it

produces an unbalance equivalent to that of a particular value of mechanical input. If this Equivalent

Input value is accurately known, it can be used as a reference point for shunt calibration of the

system. Upon completion of installation of the transducer and its associated cabling, the user can:

1. 

Perform an overall dead weight calibration using a precisely known 

value of mechanical input. The calibration can then be transferred to the 

installed calibration resistor for convenience in subsequent checking.

2. 

Replace the installed calibration resistor with one (or an equivalent 

resistance value) supplied by the transducer manufacturer to achieve a 

precisely known Equivalent Input allowing the instrument sensitivity to 

be adjusted correctly.

3. 

Determine the Equivalent Input value for the installed calibration resis- 

tor, knowing the transducer sensitivity, and adjust the instrument sen- 

sitivity accordingly.

A one percent, 59-kilohm calibration resistor is installed in the AP5103 at the factory. The installed

resistor can usually be used even though the transducer calibration data mentions some other

resistance value. In Section 4 of this manual, the techniques described above are demonstrated. If,

however, the installed value of calibration resistor is not appropriate for the transducer and

measurement range to be used, the 59-kilohm resistor should be replaced at this time. The

calibration resistor is mounted on terminals located internally to the AP5103 conditioner's printed-

circuit board. It can be accessed by removing the instrument case (Fig. 5). 

Note: A variety of DC Load Cells and Slip Ring Torque Transducers are supplied 

with the appropriate calibration resistor integral to the transducer. When this type 

of transducer is used it will be necessary to place a short across the 59-kilohm 

resistor internal to the instrument. The transducer calibration resistor can be appropri- 

ately connected to the AP5103 calibration circuit via the transducer cabling. Refer to 

cabling diragam section of this manual in the area of CAL SENSE.

This section contains the instructions for calibrating the AP5103. Included is a functional description

of the instrument front-panel (see Figure 2). To perform calibration, proceed as follows with the

sensor or calibration standard connected.

A.

Turn power ON to the AP5103 DC Power input terminal (11 to 28 Vdc input) 

The front-panel indicator should light green to indicate the application of 

dc power. Allow 10 minutes of warmup for stabilization of transducer 

characteristics. Set the Coarse Zero and Span controls label marked position.

5 of 9

Содержание AP5103 Series

Страница 1: ...All Rights Reserved Part No 234200B DCN 10087 237 Commerce Drive Amherst NY 14228 USA Tel 716 250 1900 Fax 716 250 1909 Web stellartech com Email info stellartech com Due to the nature of technology...

Страница 2: ...tion technique which is outlined in section three The supplied calibration resistor is 59 kilohms 1 which is located internally to the AP5103 case Series AP5103 DC Strain Gage Module Access switch set...

Страница 3: ...ent damage Common Mode Rejection Ratio 60 dB 1 2 excitation Input Impedance Differential and Common Mode 10 000 M Offset user adjustable vs Temperature 30 ppm C vs Time 10 ppm month Gain Accuracy Limi...

Страница 4: ...e recommended transducer cabling would be eight wire individually shielded twisted pair wired as indicated Fig 1 Sense lines must be connected at the transducer as recommended or at the AP5103 screw t...

Страница 5: ...e transducer calibration data mentions some other resistance value In Section 4 of this manual the techniques described above are demonstrated If however the installed value of calibration resistor is...

Страница 6: ...to obtain your nominal full scale analog output Use Fine Span control to adjust the signal to the precise value required or for an analog value that correspondes with the appropriate Shunt Calibration...

Страница 7: ...mote Calibration Check The instrument can be placed in the calibration mode positive or negative by shorting pin Power Common and SHP for positive or SHN for negative on the lower I O connector Figure...

Страница 8: ...is determined by the direction of the loading or off zero reading C The inability to balance correctly where the instrument output reads totally off scale and the BALANCE controls have no authority c...

Страница 9: ...ance will be made for any expenses incurred for correcting any defective workmanship and or material without written consent by Stellar Technology This warranty is in lieu of all other warranties expr...

Отзывы: