P/N 9-500-0255
3-9
Service Manual
Chapter 3: Functional Checks/Setting Defaults
The Affinity Series Control Panel
3.1
Control Panel Screens
The Affinity Series Control Panel displays three user mode screens, one debug screen and
three calibration screens. Pressing the
Reset
+
Right Arrow
or
Reset
+
Left Arrow
keys will
move to the next or previous screen. The screens are briefly described as follows:
The Run Mode Screen (Figure 3-7) is
displayed after power up. The data
fields allow the operator to select
exposure techniques, receive mes-
sages regarding system status, make
x-ray exposures, initiate compression
release and select filter. The screen
also offers a field to choose the com-
pression release method.
The User Default Screen (Figure 3-8)
allows the user to set the defaults for
compression release, compression
force, cassette sense, film ID, force
units, date, and time. The choices
made in the User Default Screen will
be displayed after power up.
Figure 3-7: The Run Mode Screen
Figure 3-8: The User Default Screen